. 2021 Apr;27(4):636.e1-636.e4.
doi: 10.1016/j.cmi.2020.12.022. Epub 2021 Jan 6.

Evaluation of a rapid antigen test (Panbio™ COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patients

Affiliations

Evaluation of a rapid antigen test (Panbio™ COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patients

Ignacio Torres et al. Clin Microbiol Infect. 2021 Apr.

Abstract

Objectives: There is limited information on the performance of rapid antigen detection (RAD) tests to identify SARS-CoV-2-infected asymptomatic individuals. In this field study, we evaluated the Panbio™ COVID-19 Ag Rapid Test Device (Abbott Diagnostics, Jena, Germany) for this purpose.

Methods: A total of 634 individuals (355 female; median age, 37 years; range, 9-87) were enrolled. Two nasopharyngeal swabs were collected from household (n = 338) and non-household contacts (n = 296) of COVID-19 cases. RAD testing was carried out at the point of care. The RT-PCR test used was the TaqPath COVID-19 Combo Kit (Thermo Fisher Scientific, MA, USA).

Results: Household contacts were tested at a median of 2 days (range, 1-7) after diagnosis of the index case, whereas non-household contacts (n = 296) were tested at a median of 6 days (range, 1-7) after exposure. In total, 79 individuals (12.4%) tested positive by RT-PCR, of whom 38 (48.1%) yielded positive RAD results. The overall sensitivity and specificity of the RAD test was 48.1% (95% CI 37.4-58.9) and 100% (95% CI 99.3-100), respectively. Sensitivity was higher in household (50.8%; 95% CI 38.9-62.5) than in non-household (35.7%; 95% CI 16.3-61.2%) contacts. Individuals testing positive by RAD test were more likely (p < 0.001) to become symptomatic than their negative counterparts.

Discussion: The Panbio test displays low sensitivity in asymptomatic close contacts of COVID-19 patients, particularly in non-household contacts. Nonetheless, establishing the optimal timing for upper respiratory tract collection in this group seems imperative to pinpoint test sensitivity.

Keywords: Asymptomatic; COVID-19; Close contacts; Rapid antigen detection test; SARS-CoV-2.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
RT-PCR cycle thresholds (Ct) (A) and SARS-CoV-2 RNA load (B) in asymptomatic close contacts of COVID-19 patients testing either positive or negative by Panbio™ COVID-19 Ag Rapid Test Device (RAD). The AMPLIRUN® TOTAL SARS-CoV-2 Control (Vircell S.A., Granada, Spain) was used as the reference material for SARS-CoV-2 RNA load quantitation (in copies/mL, considering RT-PCR Cts for the N gene). p Values for comparisons are shown.

References

    1. Options for the use of rapid antigen tests for COVID-19 in the EU/EEA and the UK. https://www.ecdc.europa.eu/en/publications-data/options-use-rapid-antige...
    1. Pekosz A., Cooper C., Parvu V., Li M., Andrews J., Manabe Y.C. Antigen-based testing but not real-time PCR correlates with SARS-CoV-2 virus culture. medRxiv. 2020 doi: 10.1101/2020.10.02.20205708. - DOI
    1. Albert E., Torres I., Bueno F., Huntley D., Molla E., Fernández-Fuentes M.Á. Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centers. Clin Microbiol Infect. 2021;27:472.e7–472.e10. - PMC - PubMed
    1. Linares M., Pérez-Tanoira R., Romanyk J., Pérez-García F., Gómez-Herruz P., Arroyo T. Panbio antigen rapid test is reliable to diagnose SARS-CoV-2 infection in the first 7 days after the onset of symptoms. J Clin Virol. 2020;133:104659. - PMC - PubMed
    1. Fenollar F., Bouam A., Ballouche M., Fuster L., Prudent E., Colson P. Evaluation of the Panbio Covid-19 rapid antigen detection test device for the screening of patients with Covid-19. J Clin Microbiol. 2020 JCM.02589-20. - PMC - PubMed

Publication types

MeSH terms

Substances