Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Mar 15;20(4):373-382.
doi: 10.1093/ons/opaa407.

Electromagnetic Navigation Systems and Intraoperative Neuromonitoring: Reliability and Feasibility Study

Affiliations

Electromagnetic Navigation Systems and Intraoperative Neuromonitoring: Reliability and Feasibility Study

Hannah Keeble et al. Oper Neurosurg. .

Abstract

Background: A recent influx of intraoperative technology is being used in neurosurgery, but few reports investigate the accuracy and safety of these technologies when used simultaneously.

Objective: To assess the ability to use an electromagnetic navigation system alongside multimodal intraoperative neurophysiological monitoring (IONM).

Methods: Single-institution prospective cohort study of patients requiring craniotomy for brain tumor resection operated using an electromagnetic navigation system (AxiEM, Medtronic®). motor evoked potentials, somatosensory evoked potentials (SSEPs), electroencephalography, and electromyography were recorded and analyzed with AxiEM on (with/without filters) and off. The neurological outcomes of the patients were recorded.

Results: A total of 15 patients were included (8 males/7 females, mean age 52.13 yr). Even though the raw acquisition is affected by the electromagnetic field (particularly SSEPs), no significant difference was detected in the morphology, amplitude, and latency of the different monitoring modalities (AxiEM off vs on) after the appropriate software filter application. Adjustments to the frequency of SSEP stimulation and number of averages, and reductions to the low-pass filters were applied. Notch filters were used appropriately and changes to the physical setup of the IONM and electromagnetic navigation system equipment reduced noise. Postoperatively, none of the patients developed new focal deficits; 7 patients showed improvement in their motor deficit (4 recovered fully).

Conclusion: The information provided by the IONM in intracranial neurosurgery patients whilst also using electromagnetic navigation systems is reliable for monitoring, mapping, and detecting intraoperative complications, provided that the appropriate software filters and tools are applied.

Keywords: Electromagnetic; Intraoperative Neuromonitoring; Neuro-Oncology; Neuronavigation.