Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
- PMID: 33441703
- PMCID: PMC7806838
- DOI: 10.1038/s41598-020-79674-4
Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
Abstract
We assess repeatability of automatic measurements of a new anterior segment optical coherence tomographer and biometer (ANTERION) and their agreement with those provided by an anterior segment-optical coherence tomography device combined with Placido-disk corneal topography (MS-39) and a validated optical biometer (IOLMaster 500). A consecutive series of patients underwent three measurements with ANTERION and one with MS-39. A subgroup of patients underwent biometry also with IOLMaster 500. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. Paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices. Repeatability of ANTERION measurements was high, with ICC > 0.98 for all parameters except astigmatism (0.963); all parameters apart from those related to astigmatism revealed a COV < 1%. Repeatability of astigmatism improved when only eyes whose keratometric astigmatism was higher than 1.0 D were investigated. Most measurements by ANTERION and MS-39 showed good agreement. No significant differences were found between measurements by ANTERION and IOLMaster, but for corneal diameter. ANTERION revealed high repeatability of automatic measurements and good agreement with both MS-39 and IOLMaster for most parameters.
Conflict of interest statement
Dr. Hoffer licenses the registered trademark name Hoffer to ensure accurate programming of his formulas to Carl Zeiss-Meditec (IOLMasters), Haag-Streit (LenStar/EyeStar), Heidelberg Engineering (ANTERION), Oculus (Pentacam AXL), Movu (Argos), Nidek (AL-Scan), Oculus (Pentacam AXL), Tomey (OA-2000), Topcon EU/VISIA Imaging (Aladdin), Ziemer (Galilei G6) and all A-scan biometer manufacturers. Dr. Savini is a consultant to CSO and has received personal fees from Alcon, Oculus and Zeiss. The remaining authors have no financial disclosures.
Similar articles
-
Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.Graefes Arch Clin Exp Ophthalmol. 2022 Jul;260(7):2271-2281. doi: 10.1007/s00417-022-05579-9. Epub 2022 Feb 16. Graefes Arch Clin Exp Ophthalmol. 2022. PMID: 35171331
-
Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus.J Refract Surg. 2024 Jul;40(7):e445-e452. doi: 10.3928/1081597X-20240514-03. Epub 2024 Jul 1. J Refract Surg. 2024. PMID: 39007818
-
Comparison of anterior segment measurements obtained using a swept-source optical coherence tomography biometer and a Scheimpflug-Placido tomographer.J Cataract Refract Surg. 2019 Mar;45(3):298-304. doi: 10.1016/j.jcrs.2018.10.033. J Cataract Refract Surg. 2019. PMID: 30851806
-
Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices.Clin Exp Ophthalmol. 2024 Nov;52(8):800-810. doi: 10.1111/ceo.14423. Epub 2024 Jul 21. Clin Exp Ophthalmol. 2024. PMID: 39034272
-
Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras.J Cataract Refract Surg. 2018 Apr;44(4):471-478. doi: 10.1016/j.jcrs.2018.02.015. Epub 2018 Apr 25. J Cataract Refract Surg. 2018. PMID: 29705008
Cited by
-
Development of an implantable collamer lens sizing model: a retrospective study using ANTERION swept-source optical coherence tomography and a literature review.BMC Ophthalmol. 2023 Feb 10;23(1):59. doi: 10.1186/s12886-023-02814-7. BMC Ophthalmol. 2023. PMID: 36765328 Free PMC article. Review.
-
Anterior-Segment Swept-Source Ocular Coherence Tomography and Scheimpflug Imaging Agreement for Keratometry and Pupil Measurements in Healthy Eyes.J Clin Med. 2021 Dec 10;10(24):5789. doi: 10.3390/jcm10245789. J Clin Med. 2021. PMID: 34945085 Free PMC article.
-
Technical failure rates for biometry between swept-source and older-generation optical coherence methods: a review and meta-analysis.BMC Ophthalmol. 2023 Apr 26;23(1):182. doi: 10.1186/s12886-023-02926-0. BMC Ophthalmol. 2023. PMID: 37101115 Free PMC article. Review.
-
Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.Graefes Arch Clin Exp Ophthalmol. 2022 Jul;260(7):2271-2281. doi: 10.1007/s00417-022-05579-9. Epub 2022 Feb 16. Graefes Arch Clin Exp Ophthalmol. 2022. PMID: 35171331
-
Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices.PLoS One. 2022 Mar 21;17(3):e0265844. doi: 10.1371/journal.pone.0265844. eCollection 2022. PLoS One. 2022. PMID: 35312733 Free PMC article.
References
-
- Savini G, Schiano-Lomoriello D, Hoffer KJ. Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras. J Cataract Refract Surg. 2018;44:471–478. doi: 10.1016/j.jcrs.2018.02.015. - DOI - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources