. 2021 Jul;31(7):5324-5334.
doi: 10.1007/s00330-020-07671-2. Epub 2021 Jan 15.

Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data

Affiliations

Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data

J Greffier et al. Eur Radiol. 2021 Jul.

Abstract

Objectives: To compare the spectral performance of dual-energy CT (DECT) platforms using task-based image quality assessment based on phantom data.

Materials and methods: Two CT phantoms were scanned on four DECT platforms: fast kV-switching CT (KVSCT), split filter CT (SFCT), dual-source CT (DSCT), and dual-layer CT (DLCT). Acquisitions on each phantom were performed using classical parameters of abdomen-pelvic examination and a CTDIvol at 10 mGy. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated from 40 to 140 keV of virtual monoenergetic images. A detectability index (d') was computed to model the detection task of two contrast-enhanced lesions as function of keV.

Results: The noise magnitude decreased from 40 to 70 keV for all DECT platforms, and the highest noise magnitude values were found for KVSCT and SFCT and the lowest for DSCT and DLCT. The average NPS spatial frequency shifted towards lower frequencies as the energy level increased for all DECT platforms, smoothing the image texture. TTF values decreased with the increase of keV deteriorating the spatial resolution. For both simulated lesions, higher detectability (d' value) was obtained at 40 keV for DLCT, DSCT, and SFCT but at 70 keV for KVSCT. The detectability of both simulated lesions was highest for DLCT and DSCT.

Conclusion: Highest detectability was found for DLCT for the lowest energy levels. The task-based image quality assessment used for the first time for DECT acquisitions showed the benefit of using low keV for the detection of contrast-enhanced lesions.

Key points: • Detectability of both simulated contrast-enhanced lesions was higher for dual-layer CT for the lowest energy levels. • The image noise increased and the image texture changed for the lowest energy levels. • The detectability of both simulated contrast-enhanced lesions was highest at 40 keV for all dual-energy CT platforms except for fast kV-switching platform.

Keywords: Comparative study; Phantoms, imaging; Radiography, dual-energy scanned projection; Tomography, X-ray computed.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Agrawal MD, Pinho DF, Kulkarni NM, Hahn PF, Guimaraes AR, Sahani DV (2014) Oncologic applications of dual-energy CT in the abdomen. Radiographics 34:589–612 - DOI
    1. De Cecco CN, Boll DT, Bolus DN et al (2017) White paper of the society of computed body tomography and magnetic resonance on dual-energy CT, part 4: abdominal and pelvic applications. J Comput Assist Tomogr 41:8–14 - DOI
    1. De Cecco CN, Schoepf UJ, Steinbach L et al (2017) White paper of the society of computed body tomography and magnetic resonance on dual-energy CT, part 3: vascular, cardiac, pulmonary, and musculoskeletal applications. J Comput Assist Tomogr 41:1–7 - DOI
    1. Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268 - DOI
    1. Goo HW, Goo JM (2017) Dual-energy CT: new horizon in medical imaging. Korean J Radiol 18:555–569 - DOI

LinkOut - more resources