Dose-response of deformable radiochromic dosimeters for spot scanning proton therapy
- PMID: 33458356
- PMCID: PMC7807645
- DOI: 10.1016/j.phro.2020.11.004
Dose-response of deformable radiochromic dosimeters for spot scanning proton therapy
Abstract
Intrafractional motion and deformation influence proton therapy delivery for tumours in the thorax, abdomen and pelvis. This study aimed to test the dose-response of a compressively strained three-dimensional silicone-based radiochromic dosimeter during proton beam delivery. The dosimeter was read-out in its relaxed state using optical computed tomography and calibrated for the linear energy transfer, based on Monte Carlo simulations. A three-dimensional gamma analysis showed a 99.3% pass rate for 3%/3 mm and 93.9% for 2%/2 mm, for five superimposed measurements using deformation-including Monte Carlo dose calculations as reference. We conclude that the dosimeter's dose-response is unaffected by deformations.
Keywords: 3D dosimetry; Deformation; Monte Carlo; Proton therapy.
© 2020 The Authors.
Conflict of interest statement
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Figures
References
-
- Chang J.Y., Zhang X., Knopf A., Li H., Mori S., Dong L. Consensus Guidelines for Implementing Pencil-Beam Scanning Proton Therapy for Thoracic Malignancies on Behalf of the PTCOG Thoracic and Lymphoma Subcommittee. Int J Radiat Oncol Biol Phys. 2017;99(1):41–50. doi: 10.1016/j.ijrobp.2017.05.014. - DOI - PubMed