Incorporation of Whipstitch Suture in Tibial Interference Fixation Improves Pullout in Anterior Cruciate Ligament Soft Tissue Grafts
- PMID: 33459342
- DOI: 10.1093/milmed/usab001
Incorporation of Whipstitch Suture in Tibial Interference Fixation Improves Pullout in Anterior Cruciate Ligament Soft Tissue Grafts
Abstract
Introduction: Interference screw fixation of soft tissue grafts is commonly used in anterior cruciate ligament (ACL) reconstruction. The purpose of this study was to determine whether including suture material at the graft-screw interface affects ultimate fixation strength of soft tissue grafts using a tibialis anterior tendon allograft model.
Materials and methods: Forty fresh-frozen human tibialis anterior tendon allografts were fixed to rigid polyurethane foam simulating the tibial tunnel. Twenty grafts underwent fixation with interference screws and 20 with interference bolts. Within each group, 10 grafts had suture in contact with either the screw or bolt. A load-to-failure test was then performed at a rate of 200 mm/min.
Results: The group of allografts with sutures in the tibial tunnel had significantly higher load to failure than the group without sutures. Using interference screw fixation, failure load of the grafts without sutures in the tunnel (535.2 ± 73.40 N) was significantly lower (P = .001) than with sutures in the tunnel (696.3 ± 110.0 N). Using interference bolt fixation, failure load of the grafts without sutures in the tunnel (613.0 ± 83.46 N) was significantly lower (P <.0001) than with sutures in the tunnel (845.8 ± 87.23 N).
Conclusions: In a biomechanical model, suture within the tibial tunnel enhances fixation strength with both interference screw and bolt fixation for soft tissue tibialis anterior allografts. Additionally, there was no difference in load to failure when comparing failure of a screw with suture in the tunnel with an interference bolt without suture. Due to improved biomechanical properties, incorporation of suture in the bone-graft interface should be considered when performing soft tissue ACL allograft reconstructions. Failure at the tibial bone-graft interface is a known complication of ACL reconstruction, and incorporation of suture within the interface should be considered for improved biomechanical properties.
Published by Oxford University Press on behalf of the Association of Military Surgeons of the United States 2021. This work is written by (a) US Government employee(s) and is in the public domain in the US.
Similar articles
-
Biomechanical comparison of 2 anterior cruciate ligament graft preparation techniques for tibial fixation: adjustable-length loop cortical button or interference screw.Am J Sports Med. 2015 Jun;43(6):1380-5. doi: 10.1177/0363546515574062. Epub 2015 Mar 12. Am J Sports Med. 2015. PMID: 25767269
-
Comparison of screw post fixation and free bone block interference fixation for anterior cruciate ligament soft tissue grafts: biomechanical considerations.Arthroscopy. 1996 Aug;12(4):470-3. doi: 10.1016/s0749-8063(96)90042-0. Arthroscopy. 1996. PMID: 8864006
-
A comparison between a retrograde interference screw, suture button, and combined fixation on the tibial side in an all-inside anterior cruciate ligament reconstruction: a biomechanical study in a porcine model.Am J Sports Med. 2009 Jan;37(1):160-7. doi: 10.1177/0363546508323747. Epub 2008 Oct 1. Am J Sports Med. 2009. PMID: 18832342
-
Hybrid Tibia Fixation of Soft Tissue Grafts in Anterior Cruciate Ligament Reconstruction: A Systematic Review.Am J Sports Med. 2016 Oct;44(10):2724-2732. doi: 10.1177/0363546515621541. Epub 2016 Jan 22. Am J Sports Med. 2016. PMID: 26801921
-
Biomechanical comparison of anterior cruciate ligament reconstruction fixation methods and implications on clinical outcomes.Ann Jt. 2023 Apr 20;8:15. doi: 10.21037/aoj-22-52. eCollection 2023. Ann Jt. 2023. PMID: 38529220 Free PMC article. Review.
Cited by
-
Mechanical properties of a bioabsorbable magnesium interference screw for anterior cruciate ligament reconstruction in various testing bone materials.Sci Rep. 2023 Jul 31;13(1):12342. doi: 10.1038/s41598-023-39513-8. Sci Rep. 2023. PMID: 37524904 Free PMC article.
-
Categorize the existing clamps used for tensile test of human graft- a systematic review.BMC Musculoskelet Disord. 2022 Jul 25;23(1):707. doi: 10.1186/s12891-022-05650-w. BMC Musculoskelet Disord. 2022. PMID: 35879684 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources