An Evolving Clinical Need: Discordant Oxygenation Measurements of Intubated COVID-19 Patients
- PMID: 33469819
- PMCID: PMC7815279
- DOI: 10.1007/s10439-020-02722-7
An Evolving Clinical Need: Discordant Oxygenation Measurements of Intubated COVID-19 Patients
Abstract
Since the first appearance of the severe acute respiratory syndrome corona virus 2 (SARS-CoV-2) earlier this year, clinicians and researchers alike have been faced with dynamic, daily challenges of recognizing, understanding, and treating the coronavirus disease 2019 (COVID-19) due to SARS-CoV-2. Those who are moderately to severely ill with COVID-19 are likely to develop acute hypoxemic respiratory failure and require administration of supplemental oxygen. Assessing the need to initiate or titrate oxygen therapy is largely dependent on evaluating the patient's existing blood oxygenation status, either by direct arterial blood sampling or by transcutaneous arterial oxygen saturation monitoring, also referred to as pulse oximetry. While the sampling of arterial blood for measurement of dissolved gases provides a direct measurement, it is technically challenging to obtain, is painful to the patient, and can be time and resource intensive. Pulse oximetry allows for non-invasive, real-time, continuous monitoring of the percent of hemoglobin molecules that are saturated with oxygen, and usually closely predicts the arterial oxygen content. As such, it was particularly concerning when patients with severe COVID-19 requiring endotracheal intubation and mechanical ventilation within one of our intensive care units were observed to have significant discordance between their predicted arterial oxygen content via pulse oximetry and their actual measured oxygen content. We offer these preliminary observations along with our speculative causes as a timely, urgent clinical need. In the setting of a COVID-19 intensive care unit, entering a patient room to obtain a fresh arterial blood gas sample not only takes exponentially longer to do given the time required for donning and doffing of personal protective equipment (PPE), it involves the consumption of already sparce PPE, and it increases the risk of viral exposure to the nurse, physician, or respiratory therapist entering the room to obtain the sample. As such, technology similar to pulse oximetry which can be applied to a patients finger, and then continuously monitored from outside the room is essential in preventing a particularly dangerous situation of unrealized hypoxia in this critically-ill patient population. Additionally, it would appear that conventional two-wavelength pulse oximetry may not accurately predict the arterial oxygen content of blood in these patients. This discordance of oxygenation measurements poses a critical concern in the evaluation and management of the acute hypoxemic respiratory failure seen in patients with COVID-19.
Keywords: Arterial blood gas; COVID-19; Hemoglobin; Oxygen saturation; Pulse oximetry; SARS-CoV-2.
Figures
Similar articles
-
Comparative Analysis of Oxygen Saturation by Pulse Oximetry and Arterial Blood Gas in Hypoxemic Patients in a Tertiary Care Hospital.Cureus. 2023 Jul 25;15(7):e42447. doi: 10.7759/cureus.42447. eCollection 2023 Jul. Cureus. 2023. PMID: 37637606 Free PMC article.
-
Advanced Uses of Pulse Oximetry for Monitoring Mechanically Ventilated Patients.Anesth Analg. 2017 Jan;124(1):62-71. doi: 10.1213/ANE.0000000000001283. Anesth Analg. 2017. PMID: 27183375 Review.
-
Correlation of oxygen saturation as measured by pulse oximetry/fraction of inspired oxygen ratio with Pao2/fraction of inspired oxygen ratio in a heterogeneous sample of critically ill children.J Crit Care. 2013 Aug;28(4):538.e1-7. doi: 10.1016/j.jcrc.2012.12.006. Epub 2013 Feb 8. J Crit Care. 2013. PMID: 23395312
-
COVID-19: Pulse oximeters in the spotlight.J Clin Monit Comput. 2021 Feb;35(1):11-14. doi: 10.1007/s10877-020-00550-7. Epub 2020 Jun 23. J Clin Monit Comput. 2021. PMID: 32578070 Free PMC article.
-
The oxygen reserve index (ORI): a new tool to monitor oxygen therapy.J Clin Monit Comput. 2018 Jun;32(3):379-389. doi: 10.1007/s10877-017-0049-4. Epub 2017 Aug 8. J Clin Monit Comput. 2018. PMID: 28791567 Free PMC article. Review.
Cited by
-
The potential role of ischaemia-reperfusion injury in chronic, relapsing diseases such as rheumatoid arthritis, Long COVID, and ME/CFS: evidence, mechanisms, and therapeutic implications.Biochem J. 2022 Aug 31;479(16):1653-1708. doi: 10.1042/BCJ20220154. Biochem J. 2022. PMID: 36043493 Free PMC article. Review.
-
Publications About COVID-19 Research by the BME Community.Ann Biomed Eng. 2022 Dec;50(12):1701-1703. doi: 10.1007/s10439-022-03068-y. Epub 2022 Sep 6. Ann Biomed Eng. 2022. PMID: 36066782 Free PMC article. No abstract available.
-
Agreement between pulse oximetry and arterial oxygen saturation measurement in critical care patients during COVID-19: a cross-sectional study.J Clin Monit Comput. 2023 Aug;37(4):977-984. doi: 10.1007/s10877-022-00959-2. Epub 2023 Jan 21. J Clin Monit Comput. 2023. PMID: 36670340 Free PMC article.
-
Characteristics of pulse oximetry and arterial blood gas in patients with fibrotic interstitial lung disease.BMJ Open Respir Res. 2024 Mar 13;11(1):e002250. doi: 10.1136/bmjresp-2023-002250. BMJ Open Respir Res. 2024. PMID: 38479819 Free PMC article.
References
-
- Enderle JD, Blanchard SM, Bronzion JD. Introduction to Biomedical Engineering. 2. San Diego: Elsevier Academic Press; 2005.
-
- Liu, W., and H. Li. Covid-19: attacks the 1-beta chain of hemoglobin and captures the porphyrin to inhibit human heme metabolism. Preprint Revised On. 10(04), 2010.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous