NeoBeat offers rapid newborn heart rate assessment
Erratum in
-
Correction: NeoBeat offers rapid newborn heart rate assessment.Arch Dis Child Fetal Neonatal Ed. 2021 Nov;106(6):e4. doi: 10.1136/archdischild-2020-320901corr1. Arch Dis Child Fetal Neonatal Ed. 2021. PMID: 34670782 No abstract available.
Abstract
Background: Heart rate (HR) is used to guide interventions during delivery room (DR) neonatal resuscitation. Dry electrode ECG (NeoBeat) may detect HR more rapidly than pulse oximetry (PO) and portable ECG, but real-time comparisons of these devices are lacking.
Design/methods: PO, ECG and NeoBeat were placed sequentially on newborns in the DR. Time for device placement and time to accurate HR acquisition were noted.
Results: DR resuscitations of 28 preterm/term infants were observed. The NeoBeat was placed faster (ie, 3 s) than PO (20 s, p=<0.0001) and ECG (16 s, p=<0.0001). Total time from initiation of device placement to HR acquisition was fastest with NeoBeat (13 s) versus ECG (42 s, p<0.0001) and PO (105 s, p<0.0001) (duration values=median).
Conclusions: These observations in a small cohort of relatively well neonates demonstrate that the NeoBeat is significantly faster to place and consistently acquires HR faster than PO and ECG.
Keywords: neonatology; resuscitation.
© Author(s) (or their employer(s)) 2021. No commercial re-use. See rights and permissions. Published by BMJ.