Is breast ultrasound a good alternative to magnetic resonance imaging for evaluating implant integrity?
- PMID: 33483142
- DOI: 10.1016/j.rx.2020.10.012
Is breast ultrasound a good alternative to magnetic resonance imaging for evaluating implant integrity?
Abstract
Objective: To compare the diagnostic performance of breast ultrasound and breast magnetic resonance imaging and to correlate the findings on the two techniques that are suggestive of implant rupture.
Material and methods: We reviewed the images and reports of breast ultrasound and magnetic resonance imaging studies done in our diagnostic and interventional breast imaging unit to evaluate breast implants in 205 patients between January 2015 and December 2017.
Results: Ultrasound findings were compatible with implant rupture in 87 (42.4%) patients: intracapsular rupture in 44 (21.5%) and intracapsular and extracapsular rupture in 43 (21.0%). Ultrasound yielded 85.2% sensitivity, 89.7% specificity, 86.2% positive predictive value, and 89.0% negative predictive value. Magnetic resonance imaging findings were compatible with implant rupture in 88 (42.9%) patients: intracapsular rupture in 50 (24.4%) and intracapsular and extracapsular rupture in 38 (18.5%). The correlation between positive findings for the location of the rupture on the two imaging techniques was excellent (0.77; p<0.0001).
Conclusion: We found high concordance between the two techniques for the detection of intracapsular and extracapsular implant rupture. These results consolidate the use of ultrasound as the first-line imaging technique to evaluate implant integrity in our population; magnetic resonance imaging can be reserved for cases in which the ultrasound diagnosis of implant integrity is uncertain.
Keywords: Breast implants; Breast/ Magnetic resonance imaging; Implantes mamarios; Mamario/ Resonancia magnética; Ultrasonido; Ultrasonography.
Copyright © 2020 SERAM. Publicado por Elsevier España, S.L.U. All rights reserved.
Similar articles
-
Is breast ultrasound a good alternative to magnetic resonance imaging for evaluating implant integrity?Radiologia (Engl Ed). 2022 Mar;64 Suppl 1:20-27. doi: 10.1016/j.rxeng.2020.10.011. Radiologia (Engl Ed). 2022. PMID: 35428463 Review.
-
Multimodality Imaging-based Evaluation of Single-Lumen Silicone Breast Implants for Rupture.Radiographics. 2017 Mar-Apr;37(2):366-382. doi: 10.1148/rg.2017160086. Epub 2017 Feb 10. Radiographics. 2017. PMID: 28186859 Review.
-
Accuracy of mammography, sonography and magnetic resonance imaging for detecting silicone breast implant ruptures: A retrospective observational study of 367 cases.Ann Chir Plast Esthet. 2021 Feb;66(1):25-41. doi: 10.1016/j.anplas.2020.09.001. Epub 2020 Sep 26. Ann Chir Plast Esthet. 2021. PMID: 32988663
-
Magnetic resonance imaging of Trilucent TM breast implants.Clin Radiol. 2002 Apr;57(4):263-7. doi: 10.1053/crad.2001.0767. Clin Radiol. 2002. PMID: 12014870
-
Imaging of breast implant and implant-associated complications: Capsular contracture and intra- or extracapsular rupture.Clin Hemorheol Microcirc. 2020;76(2):221-231. doi: 10.3233/CH-209218. Clin Hemorheol Microcirc. 2020. PMID: 32925022
LinkOut - more resources
Full Text Sources
Other Literature Sources