A new way to use transit-time flow measurement for coronary artery bypass grafting
- PMID: 33484126
- PMCID: PMC8691568
- DOI: 10.1093/icvts/ivaa328
A new way to use transit-time flow measurement for coronary artery bypass grafting
Abstract
Objectives: Transit-time flow measurement is a recognized method for graft evaluation in coronary surgery. However, single flow measurement has been associated with a low specificity for detecting graft dysfunction. The goal of this study was to assess the value of transit-time flow measurement for assessing in situ internal mammary artery grafts during non-existent native coronary circulation and the relevance of collateral blood flow in target vessels.
Methods: Between 2014 and 2018, a total of 134 patients undergoing on-pump coronary artery bypass grafting were evaluated using transit-time flow measurement. We analysed 111 single left internal mammary artery and 57 single right internal mammary artery bypasses. Correlations between coronary relevant parameters were calculated using Spearman's ρ coefficient. Risk factors for decreased flow with an arrested heart (FAH) <30 ml/min and an increased pulsatility index (PI) >3.0 as well as flow reduction >30% were calculated.
Results: FAH correlated with the diameter of the target vessel (Spearman's ρ = 0.32; P < 0.001), the amount of blood distribution (Spearman's ρ = 0.34; P < 0.001), the PI (Spearman's ρ = 0.19; P = 0.019) and the degree of stenosis (Spearman's ρ = -0.17; P = 0.042). The percentage of flow change was found to correlate with the PI (Spearman's ρ = -0.47; P < 0.0001), the degree of stenosis (Spearman's ρ = 0.42; P < 0.001), the diameter of the target vessel (Spearman's ρ = -0.22; P = 0.008) and the area of blood distribution (Spearman's ρ = -0.19; P = 0.018). A small blood distribution area was the only risk factor for decreased FAH [odds ratio (OR) 8.43, confidence interval (CI) 95% (3.04-23.41); P < 0.001]. Binary logistic regression identified PI [OR 2.05, CI 95% (1.36-3.10); P = 0.001], FAH [OR 0.98, CI 95% (0.97-0.99); P = 0.005] and degree of stenosis [OR 0.95, CI 95% (0.92-0.99); P = 0.011] as risk factors for decreased flow after cardiopulmonary bypass (<30 ml/min). An increased PI (>3) was mainly influenced by percentage of flow change [OR 0.99, CI 95% (0.98-1.00); P = 0.031].
Conclusions: FAH and percentage of flow change are related to the dimensions of the target vessel and the degree of stenosis. The addition of flow measurements with the heart arrested provides additional information about the bypass graft, the quality of the anastomosis and the physiology of the coronary circulation.
Keywords: Coronary artery bypass grafting; Quality control; Transit-time flow measurement.
© The Author(s) 2021. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
Figures


References
-
- Kolh P, Windecker S, Alfonso F, Collet JP, Cremer J, Falk V. et al. 2014 ESC/EACTS Guidelines on myocardial revascularization: the Task Force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Developed with the special contribution of the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur J Cardiothorac Surg 2014;46:517–92. - PubMed
-
- Magee MJ, Alexander JH, Hafley G, Ferguson TB Jr, Gibson CM, Harrington RA. et al. Coronary artery bypass graft failure after on-pump and off-pump coronary artery bypass: findings from PREVENT IV. Ann Thorac Surg 2008;85:494–9; discussion 499–500. - PubMed
-
- Mack MJ. Intraoperative coronary graft assessment. Curr Opin Cardiol 2008;23:568–72. - PubMed
-
- Balacumaraswami L, Abu-Omar Y, Choudhary B, Pigott D, Taggart DP.. A comparison of transit-time flowmetry and intraoperative fluorescence imaging for assessing coronary artery bypass graft patency. J Thorac Cardiovasc Surg 2005;130:315–20. - PubMed
-
- Di Giammarco G, Pano M, Cirmeni S, Pelini P, Vitolla G, Di Mauro M.. Predictive value of intraoperative transit-time flow measurement for short-term graft patency in coronary surgery. J Thorac Cardiovasc Surg 2006;132:468–74. - PubMed