. 2021 May;235(5):515-522.
doi: 10.1177/0954411921991269. Epub 2021 Jan 31.

Total knee replacement wear during simulated gait with mechanical and anatomic alignments

Affiliations

Total knee replacement wear during simulated gait with mechanical and anatomic alignments

Chase Maag et al. Proc Inst Mech Eng H. 2021 May.

Abstract

Total knee replacements (TKR) have historically been implanted perpendicular to the mechanical axis of the knee joint, with a commensurate external rotation of the femur in flexion relative to the posterior condylar axis (PCA). Although this mechanical alignment (MA) method has typically offered good long-term survivorship of implants, it may result in alignment of the implant that departs significantly from the native Joint Line (JL) in extension and flexion for a considerable portion of the patient population. There is a growing interest with surgeons to implant TKR components more closely aligned to the natural JL (Anatomic Alignment-AA) of the patient's knee joint to reduce the need for soft tissue releases during surgery, potentially improving knee function and patient satisfaction. Using a previously-validated finite element model of the lower extremity, implant- and alignment-specific loading conditions were developed and applied in a wear experiment via a six-degree-of-freedom joint simulator. MA was defined as 0° Joint Line (JL), 0° varus hip-knee-ankle (HKA) angle, and 3° external femoral rotation. AA was defined as 5° varus JL, 3° varus HKA, and 0° femoral rotation. The experiment returned wear rates of 3.76 ± 0.51 mg/million cycles (Mcyc) and 2.59 ± 2.11 mg/Mcyc for ATTUNE® cruciate-retaining (CR) fixed bearing (FB) in MA and AA, respectively. For ATTUNE posterior-stabilized (PS) FB in AA, the wear rate was 0.97 ± 1.11 mg/Mcyc. For ATTUNE CR rotating platform (RP), the wear rates were 0.23 ± 0.19 mg/Mcyc, 0.48 ± 1.02 mg/Mcyc in MA and AA respectively. Using a two-way ANOVA, it was determined that there was no significantly difference in the wear rates between AA and MA (p = 0.144) nor the wear rate of ATTUNE PS FB in AA significantly different from either ATTUNE CR FB or ATTUNE CR RP.

Keywords: Total knee replacement; experimental simulator; kinematics; load control; wear testing.