Observational Study
. 2021 Apr;114(4):287-292.
doi: 10.1016/j.acvd.2020.11.008. Epub 2021 Jan 30.

Usefulness of remote monitoring for the early detection of back-up mode in implantable cardioverter defibrillators

Affiliations
Free article
Observational Study

Usefulness of remote monitoring for the early detection of back-up mode in implantable cardioverter defibrillators

Laurence Guédon-Moreau et al. Arch Cardiovasc Dis. 2021 Apr.
Free article

Abstract

Background: Reversion of an implantable cardioverter defibrillator (ICD) to back-up mode degrades the operating capabilities of the device, puts patients at risk and requires rapid intervention by a manufacturer's technician.

Aim: To illustrate the usefulness of remote monitoring of ICDs for the early detection of reversion to back-up mode.

Methods: In our centre, all patients implanted with an ICD, with or without resynchronisation, were offered remote monitoring as soon as the technology became available. Alerts triggered by the remote monitoring system were included prospectively in a register. During a mean follow-up of 5.7±1.3 years, a total of 1594 patients with an ICD (441 with resynchronisation function) followed with remote monitoring were included in the register.

Results: Among 15,874 alerts, only 10 were related to a reversion to back-up mode. Among those, seven reversions were caused by radiotherapy, two were fake events and one was caused by magnetic resonance imaging. Except for the two fake events, the eight other patients had an emergency admission for the resetting and reprogramming of their ICD. None of the reversion to back-up mode alerts was followed by a clinical alert (i.e. a shock alert) before the ICD problem was resolved.

Conclusions: Reversion to back-up mode is a very rare event, accounting for 0.06% of total alerts; remote monitoring facilitates the early detection of this critical event to resolve the problem faster than the next scheduled follow-up. Remote monitoring can prevent serious damage to the patient and avoids systematic ambulatory control of the ICD after each radiotherapy session.

Keywords: Back-up mode; Défibrillateur automatique implantable; Implantable cardioverter defibrillator; Mode secours; Radiotherapy; Radiothérapie; Remote monitoring; Telemedicine; Télémédecine; Télésurveillance.