. 2021 Jan 29:57:8-13.
doi: 10.29390/cjrt-2020-042. eCollection 2021.

An evaluation of temperature stability and resistance in neonatal ventilator circuits

Affiliations

An evaluation of temperature stability and resistance in neonatal ventilator circuits

Jennifer A Ruppert et al. Can J Respir Ther. .

Abstract

Background: Gas conditioning minimizes complications associated with invasive ventilation of neonates. Poorly conditioned gas contributes to humidity deficit, facilitates condensate pools, and contributes to safety events. The specific aim was to objectively quantify the temperature drop across the unheated portion of a neonatal circuit and the impact condensation has to resistance to flow in the ventilator circuit.

Methods: Ventilator circuits and filters were obtained, assembled according to manufacturer recommendations, and operational verification procedures were performed prior to data collection. A neonatal test lung was connected to each Servo-I ventilator with the following settings: pressure control IMV mode; inspiratory pressure: 14 cm H2O to achieve an exhaled tidal volume of 6.0 mL; PEEP: 5 cm H2O; pressure support: 5 cm H2O, FIO2: 0.21; set frequency 40/min; and inspiratory time: 0.4 s. The Fisher and Paykel MR850 and ChonchaTherm Neptune heaters were set at a temperature of 40°C. To evaluate both systems under similar conditions, the ChonchaTherm Neptune heater humidity control was set to midline. Heaters were turned on simultaneously and given 1 h to equilibrate. Readings for room temperature, airway temperature at the patient connection, airway resistance, exhaled tidal volume, and direct observation of circuit condensation and (or) pooling were recorded hourly for a 48-h period. Summary statistics were calculated for the variables of interest.

Results: Mean (±SD) air temperature was 26.3°C (±1.4) for the Fisher & Paykel MR850 system and 26.2°C (±1.5), for the ChonchaTherm Neptune system. Mean (±SD) airway resistance was 229.3 cm H2O/L/s (±81.0) for the Fisher & Paykel system and 196.2 cm H2O/L/s (±39.4) for the ChonchaTherm Neptune system. Mean (±SD) tidal volume for the Fisher & Paykel MR850 system was 6.5 mL (±0.4), and for the ChonchaTherm Neptune system was 7.2 mL (±0.6).

Conclusion: Circuit condensate increased tidal volume delivery and airway resistance. Temperature at the patient connection was lower than the temperature monitored by the system 12 inches distally, which can negatively impact gas conditioning.

Keywords: condensate; gas conditioning; humidification; lavage; neonatal ventilation; temperature.

PubMed Disclaimer

Figures

Figure 1
Figure 1. Individual and moving range chart displaying temperatures recorded hourly in °C using the Fisher and Paykel MR 850 heating system. Temperatures displayed in red with a black indicator marker above the datapoint represent those that exceeded the upper confidence limit.
Figure 2
Figure 2. Individual and moving range chart displaying temperatures recorded hourly in °C using the ChonchaTherm Neptune heating system. Temperatures displayed in red with a black indicator marker above the datapoint represent those that exceeded the upper or lower confidence limits.
Figure 3
Figure 3. Individual and moving range chart displaying hourly airway resistance values, recorded in cmH2O/L/s, using the Fisher and Paykel MR850 heating system. Values displayed in red with a black indicator marker above the datapoint represent those that exceeded the upper or lower confidence limits.
Figure 4
Figure 4. Individual and moving range chart displaying hourly airway resistance values, recorded in cmH2O/L/s, using the ChonchaTherm Neptune heating system. Values displayed in red with a black indicator marker above the datapoint represent those that exceeded the upper confidence limit.

References

    1. Schulze A. Respiratory gas conditioning and humidification. Clin Perinatol 2007;34(1):19–33. doi: 10.1016/j.clp.2006.12.009. - DOI - PubMed
    1. Sosulski R, Polin RA, Baumgart S. Respiratory water loss and heat balance in intubated infants receiving humidified air. J Pediatr 1983;103(2):307–10. doi: 10.1016/S0022-3476(83)80374-6. - DOI - PubMed
    1. Plotnikow GA, Accoce M, Navarro E, Tiribelli N. Humidification and heating of inhaled gas in patients with artificial airway. A narrative review. Rev Bras Ter Intensiva 2018;30(1):86–97. doi: 10.5935/0103-507X.20180015. - DOI - PMC - PubMed
    1. Tarnow-Mordi WO, Reid E, Griffiths P, Wilkinson AR. Low inspired gas temperature and respiratory complications in very low birth weight infants. J Pediatr 1989;114(3):438–42. doi: 10.1016/S0022-3476(89)80567-0. - DOI - PubMed
    1. Schulze A. Respiratory gas conditioning in infants with an artificial airway. Semin Neonatol 2002;7(5):369–77. doi: 10.1053/siny.2002.0131. - DOI - PubMed

LinkOut - more resources