. 2021 Feb 3;223(2):206-213.
doi: 10.1093/infdis/jiaa641.

Rapid, Sensitive, and Specific Severe Acute Respiratory Syndrome Coronavirus 2 Detection: A Multicenter Comparison Between Standard Quantitative Reverse-Transcriptase Polymerase Chain Reaction and CRISPR-Based DETECTR

Affiliations

Rapid, Sensitive, and Specific Severe Acute Respiratory Syndrome Coronavirus 2 Detection: A Multicenter Comparison Between Standard Quantitative Reverse-Transcriptase Polymerase Chain Reaction and CRISPR-Based DETECTR

Eelke Brandsma et al. J Infect Dis. .

Abstract

Background: Recent advances in CRISPR-based diagnostics suggest that DETECTR, a combination of reverse-transcriptase loop-mediated isothermal amplification (RT-LAMP) and subsequent Cas12 bystander nuclease activation by amplicon-targeting ribonucleoprotein complexes, could be a faster and cheaper alternative to quantitative reverse-transcription polymerase chain reaction (qRT-PCR) without sacrificing sensitivity and/or specificity.

Methods: In this study, we compare DETECTR with qRT-PCR to diagnose coronavirus disease 2019 on 378 patient samples. Patient sample dilution assays suggest a higher analytical sensitivity of DETECTR compared with qRT-PCR; however, this was not confirmed in this large patient cohort, where we report 95% reproducibility between the 2 tests.

Results: These data showed that both techniques are equally sensitive in detecting severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) providing additional value of DETECTR to the currently used qRT-PCR platforms. For DETECTR, different guide ribonucleic acids can be used simultaneously to obviate negative results due to mutations in N-gene. Lateral flow strips, suitable as a point-of-care test, showed a 100% correlation to the high-throughput DETECTR assay. More importantly, DETECTR was 100% specific for SARS-CoV-2 relative to other human coronaviruses.

Conclusions: Because there is no need for specialized equipment, DETECTR could be rapidly implemented as a complementary technically independent approach to qRT-PCR thereby increasing the testing capacity of medical microbiological laboratories and relieving the existent PCR platforms for routine non-SARS-CoV-2 diagnostic testing.

Keywords: COVID-19; DETECTR; SARS-CoV-2; qRT-PCR.