Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Oct;23(10):906-920.
doi: 10.1177/1098612X20983260. Epub 2021 Feb 5.

Comparison of serum and plasma SDMA measured with point-of-care and reference laboratory analysers: implications for interpretation of SDMA in cats

Affiliations

Comparison of serum and plasma SDMA measured with point-of-care and reference laboratory analysers: implications for interpretation of SDMA in cats

Randolph M Baral et al. J Feline Med Surg. 2021 Oct.

Abstract

Objectives: Symmetric dimethylarginine (SDMA) reflects the glomerular filtration rate (GFR) in people, dogs and cats. Initial assays used a liquid chromatography-mass spectroscopy (LC-MS) technique. A veterinary immunoassay has been developed for use in commercial laboratories and point-of-care (POC) laboratory equipment. This study sought to: determine POC and commercial laboratory (CL) SDMA assay imprecision; determine any bias of the POC assay compared with the CL assay; calculate observed total error of the POC assay and compare with analytical performance goals; and calculate dispersion and sigma metrics (σ) for POC and CL SDMA methods.

Methods: Two separate studies were performed that assessed: (1) imprecision, determined by evaluation of pooled feline plasma or serum; and (2) bias, assessed by comparing pooled plasma and serum results, as well as paired analyses of clinical samples from a single venepuncture measured using both analysers. Results were assessed in relation to performance goals. Dispersion and σ were calculated for both analysers.

Results: Bias between CL and POC analysers was consistent and high numbers of clinical results were outside performance goals across both studies. Imprecision was poor for both analysers for study 1 and improved to within quality goals for the CL analyser for study 2. Dispersion was at least 40%, meaning a measured result of 14 μg/dl represents a range of possible results from 8 μg/dl to 20 μg/dl.

Conclusions and relevance: Clinicians should be careful ascribing medical significance to small changes in SDMA concentration, as these may reflect analytical and biological variability. Analyser-specific reference intervals are likely required.

Keywords: Symmetric dimethylarginine; bias; dispersion; imprecision; method comparison; quality goal; sigma metrics; test interpretation.

PubMed Disclaimer

Conflict of interest statement

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Figures

Figure 1
Figure 1
Flow chart of study 1 procedures and statistics for pooled heparinised plasma samples to investigate symmetric dimethylarginine (SDMA) and creatinine imprecision and related calculations. CVA = analyser coefficient of variation (imprecision); CVI = individual coefficient of variation (biological variation) from prior study
Figure 2
Figure 2
Flow chart of study 1 procedures and statistical assessments to compare symmetric dimethylarginine (SDMA) results from a point-of-care (POC) analyser measured using heparinised plasma with SDMA results from a commercial laboratory (CL) analyser measured using serum. Samples were from healthy cats and cats with previously diagnosed kidney disease
Figure 3
Figure 3
Flow chart of study 2 procedures and statistics for pooled heparinised plasma samples and pooled serum samples to investigate symmetric dimethylarginine (SDMA) and creatinine imprecision and related calculations. CVA = analyser coefficient of variation (imprecision); CVI = individual coefficient of variation (biological variation) from prior study
Figure 4
Figure 4
Flow chart of study 2 procedures and statistical assessments to compare SDMA results from a point-of-care (POC) analyser and a commercial laboratory (CL) analyser measured using heparinised plasma from healthy cats and cats with previously diagnosed kidney disease
Figure 5
Figure 5
Scatter plot with line of best fit of results from feline serum and plasma symmetric dimethylarginine (SDMA) concentrations analysed by a commercial laboratory (CL) analyser and point-of-care (POC) analyser prior to optimisation of the CL method (study 1). Note that the individual results are scattered around rather than sitting close to the line of best fit. Note that one very high result (65 μg/dl and 73 μg/dl on each analyser) has been excluded for clarity
Figure 6
Figure 6
Scatter plot with line of best fit of results from feline plasma symmetric dimethylarginine (SDMA) concentrations analysed by a commercial laboratory (CL) analyser and point-of-care (POC) analyser after optimisation of the CL method (study 2). Note that the individual results are scattered around rather than sitting close to the line of best fit
Figure 7
Figure 7
Bland–Altman difference plot for study 1 constructed by plotting for each cat the mean concentration from both analysers (x axis) against the difference between results from each analyser (y axis). The yellow lines indicate the allowable total error (TEA) as determined by expert opinion (via the survey). The blue sloping lines indicate the desirable total error (TED) as determined by a prior biological variation study. The purple lines indicate the limits of agreement (LOA) to 95% confidence interval. The LOA range is approximately double the two determined measures of clinical importance (TEA and TED) and are therefore too high for clinical acceptability
Figure 8
Figure 8
Bland–Altman difference plot for study 2 assessing the mean concentration of each analyser to the difference. The negative bias of the point-of-care (POC) analyser appears to increase as the concentration of symmetric dimethylarginine (SDMA) increases. The limits of agreement (LOA) are very similar to study 1 and too high for clinical acceptability. TEA = allowable total error; TED = desirable total error

Similar articles

Cited by

References

    1. Kielstein JT, Salpeter SR, Bode-Boeger SM, et al.. Symmetric dimethylarginine (SDMA) as endogenous marker of renal function – a meta-analysis. Nephrol Dial Transplant 2006; 21: 2446–2451. - PubMed
    1. Nabity M, Lees G, Boggess M, et al.. Symmetric dimethylarginine assay validation, stability, and evaluation as a marker for the early detection of chronic kidney disease in dogs. J Vet Intern Med 2015; 29: 1036–1044. - PMC - PubMed
    1. Baral RM. Assessment of renal disease in cats by SDMA, creatinine assessed by population reference interval and creatinine assessed by reference change value [abstract]. J Feline Med Surg 2016; 18: 935.
    1. Braff J, Obare E, Yerramilli M, et al.. Relationship between serum symmetric dimethylarginine concentration and glomerular filtration rate in cats. J Vet Intern Med 2014; 28: 1699–1701. - PMC - PubMed
    1. Patch D, Obare E, Prusevich P, et al.. High throughput immunoassay for kidney function biomarker symmetric dimethylarginine (SDMA) [abstract]. Clin Chem 2015; 61: S135.

LinkOut - more resources