Multicenter Study
. 2021 Jul;30(7):1905-1911.
doi: 10.1007/s00586-021-06750-2. Epub 2021 Feb 5.

Metallosis is commonly associated with magnetically controlled growing rods; results from an independent multicentre explant database

Affiliations
Multicenter Study

Metallosis is commonly associated with magnetically controlled growing rods; results from an independent multicentre explant database

Paul R P Rushton et al. Eur Spine J. 2021 Jul.

Abstract

Purpose: Determine the incidence of metallosis around MAGEC rods.

Methods: A multicentre explant database was searched to identify cases with complete intraoperative findings at rod removal. Surgeons removing rods detailed the presence or absence of tissue metallosis associated with rods. More recently surgeons measured the 'length' of tissue metallosis. Prior to rod disassembly, the majority underwent testing with an external remote controller (ERC). The impact of clinical and explant variables on metallosis was assessed.

Results: Sixty-six cases were identified. Mean age at insertion was 8.1 ± 2.3 years with mean duration of implantation 37.6 ± 15.1 months. Tissue metallosis was noted at revision surgery in 52/66 cases (79%). Metallosis was noted more commonly when rods were removed during fusion surgery than rod removal/exchange (97% vs. 58% (p = < 0.01)). The mass at insertion was greater in cases with metallosis (25.9 ± 7.8 kg vs. 21.1 ± 6.2 kg, p = 0.04). Length of tissue metallosis was reported for 45 rods, median 9 cm (range 1-25). Metallosis was noted in 43/59 (73%) rods that produced no force and 22/30 (73%) rods that produced some force on ERC activation (p = 0.96). Wear debris was found within the actuator in all rods, and all but 3 rods had damaged O-rings.

Conclusion: MAGEC rods are associated with tissue metallosis in the majority of cases. It is seen with functional rods as well as failed rods and appears related to wear debris within the actuator and high rates of O-ring failure. Until the implications of metal debris in children are known, we urge caution with the use of this implant.

Keywords: Explant analysis; Growing rods; MAGEC; Magnetically controlled; Metallosis.

PubMed Disclaimer

References

    1. Cheung KM-C, Cheung JP-Y, Samartzis D et al (2012) Magnetically controlled growing rods for severe spinal curvature in young children: a prospective case series. Lancet 379:1967–1974 - DOI
    1. Nnadi C, Thakar C, Wilson-MacDonald J et al (2018) An NIHR-approved two-year observational study on magnetically controlled growth rods in the treatment of early onset scoliosis. Bone Jt J 100-B:507–515 - DOI
    1. NuVasive Urgent Field Safety Notice / MAGEC System. 2019.
    1. MHRA MDA/2020/010 Spinal implant: MAGEC System Model X rods—risk of failure in use. 2020.
    1. MHRA MDA/2020/011 Spinal implant: All MAGEC Systems—supply suspended to the UK. 2020.

Publication types

LinkOut - more resources