Maintaining Image Quality While Reducing Acoustic Noise and Switched Gradient Field Exposure During Lumbar MRI
- PMID: 33565199
- DOI: 10.1002/jmri.27527
Maintaining Image Quality While Reducing Acoustic Noise and Switched Gradient Field Exposure During Lumbar MRI
Abstract
Background: MR-generated acoustic noise can contribute to patient discomfort and potentially be harmful. One way to reduce this noise is by altering the gradient output and/or waveform using software optimization. Such modifications might influence image quality and switched gradient field exposure, and different techniques appear to affect sound pressure levels (SPLs) to various degrees.
Purpose: To evaluate SPLs, image quality, switched gradient field exposure, and participants' perceived noise levels during two different acoustic noise reduction (ANR) techniques, Quiet Suite (QS) and Whisper Mode (WM), and to compare them with conventional T2-weighted turbo spin echo (T2W TSE) of the lumbar spine.
Design: Prospective.
Subjects: Forty adults referred for lumbar MRI.
Field strength/sequence: Conventional T2W TSE, T2W TSE with QS, and T2W TSE with WM were acquired at 1.5 T.
Assessment: Peak SPL (A-weighted decibels, dBA), perceived noise levels (Borg CR10®-scale), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), three radiologists' qualitative assessments in image quality on an ordinal scale 1-4, switched gradient field exposure (% general public), and gradient currents were measured. Interobserver reliability was reported as percentage agreement.
Statistical tests: Repeated measures ANOVA, Friedman's ANOVA, and Wilcoxon's Signed-Rank Test for acoustic noise measurements and image quality assessments.
Results: Mean peak SPLs were 89.9 dBA, 74.3 dBA, and 78.8 dBA for conventional, QS, and WM, respectively (P < 0.05). Participants perceived QS as the quietest and conventional as the loudest sequence (P < 0.05). No qualitative differences in image quality were seen (P > 0.05), although QS showed significantly improved SNR and CNR (P < 0.05). Switched gradient field exposure was reduced by 66% and 48% for QS and WM, respectively.
Data conclusion: Without degrading image quality, both QS and WM are viable ANR techniques in lumbar T2W TSE. QS provided the lowest SPL, the lowest gradient field exposure and was perceived as the most silent among the three sequences.
Level of evidence: 1 TECHNICAL EFFICACY STAGE: 5.
Keywords: MRI safety; magnetic field; patient comfort; quiet suite; whisper mode.
© 2021 The Authors. Journal of Magnetic Resonance Imaging published by Wiley Periodicals LLC. on behalf of International Society for Magnetic Resonance in Medicine.
Similar articles
-
Evaluation of Software-Optimized Protocols for Acoustic Noise Reduction During Brain MRI at 7 Tesla.J Magn Reson Imaging. 2025 Aug;62(2):577-587. doi: 10.1002/jmri.29749. Epub 2025 Mar 6. J Magn Reson Imaging. 2025. PMID: 40048635 Free PMC article.
-
Evaluation of magnetic resonance imaging acoustic noise reduction technology by magnetic gradient waveform control.Magn Reson Imaging. 2019 Nov;63:170-177. doi: 10.1016/j.mri.2019.08.015. Epub 2019 Aug 16. Magn Reson Imaging. 2019. PMID: 31425801
-
Acoustic noise reduction in T 1- and proton-density-weighted turbo spin-echo imaging.MAGMA. 2016 Feb;29(1):5-15. doi: 10.1007/s10334-015-0502-7. Epub 2015 Oct 22. MAGMA. 2016. PMID: 26490348
-
Acoustic Noise and Magnetic Resonance Imaging: A Narrative/Descriptive Review.J Magn Reson Imaging. 2022 Feb;55(2):337-346. doi: 10.1002/jmri.27525. Epub 2021 Feb 25. J Magn Reson Imaging. 2022. PMID: 33629790 Review.
-
Advancements in Gradient System Performance for Clinical and Research MRI.J Magn Reson Imaging. 2023 Jan;57(1):57-70. doi: 10.1002/jmri.28421. Epub 2022 Sep 8. J Magn Reson Imaging. 2023. PMID: 36073722 Review.
Cited by
-
Evaluation of Software-Optimized Protocols for Acoustic Noise Reduction During Brain MRI at 7 Tesla.J Magn Reson Imaging. 2025 Aug;62(2):577-587. doi: 10.1002/jmri.29749. Epub 2025 Mar 6. J Magn Reson Imaging. 2025. PMID: 40048635 Free PMC article.
-
Effect of acoustic noise reduction technology on image quality: a multivendor study.Radiol Phys Technol. 2023 Jun;16(2):235-243. doi: 10.1007/s12194-023-00712-7. Epub 2023 Mar 25. Radiol Phys Technol. 2023. PMID: 36964891
-
MRI Safety Practice Observations in MRI Facilities Within the Kingdom of Jordan, Compared to the 2020 Manual on MR Safety of the American College of Radiology.Med Devices (Auckl). 2022 May 13;15:131-142. doi: 10.2147/MDER.S360335. eCollection 2022. Med Devices (Auckl). 2022. PMID: 35592097 Free PMC article.
-
GTV delineating for patients with postoperative glioma based on enhanced T2-FLAIR sequence instead of enhanced T1-TFE sequence: a feasibility study.Discov Oncol. 2025 May 25;16(1):919. doi: 10.1007/s12672-025-02697-8. Discov Oncol. 2025. PMID: 40413710 Free PMC article.
References
-
- McJury M, Shellock FG. Auditory noise associated with MR procedures: A review. J Magn Reson Imaging 2000;12:37-45.
-
- McRobbie D, Moore E, Graves M, Prince M. MRI from picture to proton. 3rd ed. Cambridge: Cambridge University Press; 2017. p 153-154.
-
- Baker MA. Reduction of MRI acoustic noise achieved by manipulation of scan parameters - A study using veterinary MR sequences. Radiography 2013;19:11-16.
-
- Heismann B, Ott M, Grodzki D. Sequence-based acoustic noise reduction of clinical MRI scans. Magn Reson Med 2015;73:1104-1109.
-
- Corcuera-Solano I, Doshi A, Pawha PS, Gui D, Gaddipati A, Tanenbaum L. Quiet PROPELLER MRI techniques match the quality of conventional PROPELLER brain imaging techniques. AJNR Am J Neuroradiol 2015;36:1124-1127.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials