Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Feb 8;10(4):639.
doi: 10.3390/jcm10040639.

Long-Term Technical Performance of the Osypka QT-5® Ventricular Pacemaker Lead

Affiliations

Long-Term Technical Performance of the Osypka QT-5® Ventricular Pacemaker Lead

Georg Semmler et al. J Clin Med. .

Abstract

Background: Lead-associated complications and technical issues in patients with cardiac implantable electronic devices are common but underreported in the literature.

Methods: All patients undergoing implantation of the Osypka QT-5® ventricular lead at the University Clinic St. Pölten between 1 January 2006 and 31 December 2012 were retrospectively analyzed (n = 211). Clinical data including pacemaker follow-up examinations and the need for lead revisions were assessed. Kaplan-Meier analysis to estimate the rate of lead dysfunction during long-term follow-up was conducted.

Results: Patients were followed for a median of 5.2 years (interquartile range (IQR) 2.0-8.7). R-wave sensing properties at implantation, compared to last follow-up, remained basically unchanged: 9.9 mV (IQR 6.8-13.4) and 9.6 mV (IQR 5.6-12.0), respectively). Ventricular pacing threshold significantly increased between implantation (0.5 V at 0.4 ms; IQR 0.5-0.8) and the first follow-up visit (1.0 V at 0.4 ms; IQR 0.8-1.3; p < 0.001) and this increase persisted throughout to the last check-up (0.9 V at 0.4 ms; IQR 0.8-1.2). Impedance significantly declined from 1142 Ω (IQR 955-1285) at implantation to 814 Ω (IQR 701-949; p < 0.001) at the first check-up, followed by a further decrease to 450 Ω (IQR 289-652; p < 0.001) at the last check-up. Overall, the Osypka QT-5® ventricular lead was replaced in 36 patients (17.1%).

Conclusions: This report shows an unexpected high rate of technical issues of the Osypka QT-5® ventricular lead during long-term follow-up.

Keywords: impedance decline; lead replacement; pacemaker malfunction; reoperation.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Figure 1
Figure 1
Device characteristics (lead impedance, sensing and pacing threshold) at implantation and at last follow-up: (A) differences in ventricular lead impedance at implantation and at last follow-up; (B) differences in ventricular sensing at implantation and at last follow-up; (C) differences in ventricular pacing threshold at implantation and at last follow-up. Whiskers in box plots mark the 95% (2.5–97.5%) confidence interval.
Figure 2
Figure 2
Kaplan-Meier curve describing the prevalence of dysfunction of the Osypka QT-5® ventricular lead.

References

    1. Sidhu S., Marine J.E. Evaluating and managing bradycardia. Trends Cardiovasc. Med. 2019;30:265–272. doi: 10.1016/j.tcm.2019.07.001. - DOI - PubMed
    1. Langenfeld H., Grimm W., Maisch B., Koghsiek K. Course of Symptoms and Spontaneous ECG in Pacemaker Patients: A 5-Year Follow-up Study. Pacing Clin. Electrophysiol. 1988;11:2198–2206. doi: 10.1111/j.1540-8159.1988.tb05986.x. - DOI - PubMed
    1. Alboni P., Menozzi C., Brignole M., Paparella N., Gaggioli G., Lolli G., Cappato R. Effects of permanent pacemaker and oral theophylline in sick sinus syndrome the THEOPACE study: A randomized controlled trial. Circulation. 1997;96:260–266. doi: 10.1161/01.CIR.96.1.260. - DOI - PubMed
    1. Edhag O., Swahn A. Prognosis of patients with complete heart block or arrhythmic syncope who were not treated with artificial pacemakers. A long-term follow-up study of 101 patients. Acta Medica Scand. 1976;200:457–463. doi: 10.1111/j.0954-6820.1976.tb08264.x. - DOI - PubMed
    1. Udo E.O., Zuithoff N.P., Van Hemel N.M., De Cock C.C., Hendriks T., Doevendans P.A., Moons K.G. Incidence and predictors of short- and long-term complications in pacemaker therapy: The Followpace study. Hearth Rhythm. 2012;9:728–735. doi: 10.1016/j.hrthm.2011.12.014. - DOI - PubMed

LinkOut - more resources