A comparison of 8 and 16 frames gated SPECT imaging for determination of left ventricular volumes and ejection fraction: effects of gender and myocardial counts
- PMID: 33575863
- DOI: 10.1007/s10554-021-02170-x
A comparison of 8 and 16 frames gated SPECT imaging for determination of left ventricular volumes and ejection fraction: effects of gender and myocardial counts
Abstract
In myocardial gated SPECT imaging each cardiac cycle is divided into 8 or 16 temporal frames and the cause of the difference between 8 and 16 frames is not specified exactly. The aim of this study was to investigate the effect of myocardial detector counts and gender on the difference between 8 and 16 frames and also to compare the LVEF obtained by 8 and 16 frames with echocardiography. The study population included 84 patients who underwent gated SPECT imaging. Left ventricular parameters were assessed on 8 and 16 frames gated SPECT. LVEF was also measured with two-dimensional echocardiography within 5-10 days after gated SPECT imaging. There was a good correlation between 8 and 16 frames for calculation of LVEF (p = 0.00, r = 0.860), EDV (p = 0.00, r = 0.965) and ESV (p = 0.00, r = 0.956) in all patients. But the difference between 8 and 16 frames for calculation of LVEF (p = 0.00), EDV (p = 0.014) and ESV (p = 0.00) was statistically significant. This difference was assessed separately in females, males, patients with high photon counts and patients with low photon counts and in all subgroups was statistically significant difference in the estimation of LVEF and ESV (p < 0.05) but no significant difference in the estimation of EDV (p > 0.05). Echocardiography resulted in smaller LVEF as compared to 8 and 16 frames gated SPECT studies and there was a significant difference between the two methods (p = 0.00). The myocardial detector counts and gender have no effect on the difference between 8 and 16 frames methods and the LVEF on echocardiography is smaller than the gated SPECT, but the 8-frame is closer to echocardiography.
Keywords: 8 and 16 frames gated SPECT; Echocardiography; Ejection fraction; LV volumes.
Similar articles
-
Assessment of left ventricular function and volumes for patients with dilated cardiomyopathy using gated myocardial perfusion SPECT and comparison with echocardiography.Nucl Med Commun. 2005 Aug;26(8):701-10. doi: 10.1097/01.mnm.0000170938.98581.14. Nucl Med Commun. 2005. PMID: 16000988 Clinical Trial.
-
Comparison of two software in gated myocardial perfusion single photon emission tomography, for the measurement of left ventricular volumes and ejection fraction, in patients with and without perfusion defects.Hell J Nucl Med. 2007 Jan-Apr;10(1):19-23. Hell J Nucl Med. 2007. PMID: 17450245
-
Left ventricular ejection fraction and volumes on rest gated 201Tl perfusion SPECT: comparison with two-dimensional echocardiography.Nucl Med Commun. 2006 May;27(5):425-9. doi: 10.1097/00006231-200605000-00003. Nucl Med Commun. 2006. PMID: 16609353 Clinical Trial.
-
Comparison of gated SPECT, echocardiography and cardiac magnetic resonance imaging for the assessment of left ventricular ejection fraction and volumes.Ann Saudi Med. 2007 Nov-Dec;27(6):415-20. doi: 10.5144/0256-4947.2007.415. Ann Saudi Med. 2007. PMID: 18059128 Free PMC article. Clinical Trial.
-
Quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPECT: MRI validation and comparison of the Emory Cardiac Tool Box with QGS and 4D-MSPECT.J Nucl Med. 2005 Aug;46(8):1256-63. J Nucl Med. 2005. PMID: 16085580 Clinical Trial.
References
-
- Go V, Bhatt MR, Hendel RC (2004) The diagnostic and prognostic value of ECG-gated SPECT myocardial perfusion imaging. J Nucl Med 45(5):912–921 - PubMed
-
- Paul AK, Nabi HA (2004) Gated myocardial perfusion SPECT: basic principles, technical aspects, and clinical applications. J Nucl Med Technol 32(4):179–187 - PubMed
-
- Kurisu S, Sumimoto Y, Ikenaga H, Watanabe N, Ishibashi K, Dohi Y, Fukuda Y, Kihara Y (2017) Comparison of 8-frame and 16-frame thallium-201 gated myocardial perfusion SPECT for determining left ventricular systolic and diastolic parameters. Heart Vessels 32(7):790–795. https://doi.org/10.1007/s00380-016-0935-6 - DOI - PubMed
-
- Montelatici G, Sciagrà R, Passeri A, Dona M, Pupi A (2008) Is 16-frame really superior to 8-frame gated SPECT for the assessment of left ventricular volumes and ejection fraction? Comparison of two simultaneously acquired gated SPECT studies. Eur J Nucl Med Mol Imaging 35(11):2059–2065. https://doi.org/10.1007/s00259-008-0866-2 - DOI - PubMed
-
- Navare SM, Wackers FJ, Liu Y-H (2003) Comparison of 16-frame and 8-frame gated SPET imaging for determination of left ventricular volumes and ejection fraction. Eur J Nucl Med Mol Imaging 30(10):1330–1337. https://doi.org/10.1007/s00259-003-1231-0 - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources