Improved Glycemic Outcomes With Medtronic MiniMed Advanced Hybrid Closed-Loop Delivery: Results From a Randomized Crossover Trial Comparing Automated Insulin Delivery With Predictive Low Glucose Suspend in People With Type 1 Diabetes
- PMID: 33579715
- DOI: 10.2337/dc20-2250
Improved Glycemic Outcomes With Medtronic MiniMed Advanced Hybrid Closed-Loop Delivery: Results From a Randomized Crossover Trial Comparing Automated Insulin Delivery With Predictive Low Glucose Suspend in People With Type 1 Diabetes
Abstract
Objective: To study the MiniMed Advanced Hybrid Closed-Loop (AHCL) system, which includes an algorithm with individualized basal target set points, automated correction bolus function, and improved Auto Mode stability.
Research design and methods: This dual-center, randomized, open-label, two-sequence crossover study in automated-insulin-delivery-naive participants with type 1 diabetes (aged 7-80 years) compared AHCL to sensor-augmented pump therapy with predictive low glucose management (SAP + PLGM). Each study phase was 4 weeks, preceded by a 2- to 4-week run-in and separated by a 2-week washout.
Results: The study was completed by 59 of 60 people (mean age 23.3 ± 14.4 years). Time in target range (TIR) 3.9-10 mmol/L (70-180 mg/dL) favored AHCL over SAP + PLGM (70.4 ± 8.1% vs. 57.9 ± 11.7%) by 12.5 ± 8.5% (P < 0.001), with greater improvement overnight (18.8 ± 12.9%, P < 0.001). All age-groups (children [7-13 years], adolescents [14-21 years], and adults [>22 years]) demonstrated improvement, with adolescents showing the largest improvement (14.4 ± 8.4%). Mean sensor glucose (SG) at run-in was 9.3 ± 0.9 mmol/L (167 ± 16.2 mg/dL) and improved with AHCL (8.5 ± 0.7 mmol/L [153 ± 12.6 mg/dL], P < 0.001), but deteriorated during PLGM (9.5 ± 1.1 mmol/L [17 ± 19.8 mg/dL], P < 0.001). TIR was optimal when the algorithm set point was 5.6 mmol/L (100 mg/dL) compared with 6.7 mmol/L (120 mg/dL), 72.0 ± 7.9% vs. 64.6 ± 6.9%, respectively, with no additional hypoglycemia. Auto Mode was active 96.4 ± 4.0% of the time. The percentage of hypoglycemia at baseline (<3.9 mmol/L [70 mg/dL] and ≤3.0 mmol/L [54 mg/dL]) was 3.1 ± 2.1% and 0.5 ± 0.6%, respectively. During AHCL, the percentage time at <3.9 mmol/L (70 mg/dL) improved to 2.1 ± 1.4% (P = 0.034) and was statistically but not clinically reduced for ≤3.0 mmol/L (54 mg/dL) (0.5 ± 0.5%; P = 0.025). There was one episode of mild diabetic ketoacidosis attributed to an infusion set failure in combination with an intercurrent illness, which occurred during the SAP + PLGM arm.
Conclusions: AHCL with automated correction bolus demonstrated significant improvement in glucose control compared with SAP + PLGM. A lower algorithm SG set point during AHCL resulted in greater TIR, with no increase in hypoglycemia.
© 2021 by the American Diabetes Association.
Similar articles
-
Improved technology satisfaction and sleep quality with Medtronic MiniMed® Advanced Hybrid Closed-Loop delivery compared to predictive low glucose suspend in people with Type 1 Diabetes in a randomized crossover trial.Acta Diabetol. 2022 Jan;59(1):31-37. doi: 10.1007/s00592-021-01789-5. Epub 2021 Aug 27. Acta Diabetol. 2022. PMID: 34453208 Clinical Trial.
-
Rapid Improvement in Time in Range After the Implementation of an Advanced Hybrid Closed-Loop System in Adolescents and Adults with Type 1 Diabetes.Diabetes Technol Ther. 2021 Sep;23(9):609-615. doi: 10.1089/dia.2021.0037. Epub 2021 Apr 20. Diabetes Technol Ther. 2021. PMID: 33784187
-
Safety and Glycemic Outcomes During the MiniMedTM Advanced Hybrid Closed-Loop System Pivotal Trial in Children and Adolescents with Type 1 Diabetes.Diabetes Technol Ther. 2023 Nov;25(11):755-764. doi: 10.1089/dia.2023.0255. Epub 2023 Oct 25. Diabetes Technol Ther. 2023. PMID: 37782145 Clinical Trial.
-
The efficacy and safety of insulin pump therapy with predictive low glucose suspend feature in decreasing hypoglycemia in children with type 1 diabetes mellitus: A systematic review and meta-analysis.Pediatr Diabetes. 2020 Nov;21(7):1256-1267. doi: 10.1111/pedi.13088. Epub 2020 Sep 14. Pediatr Diabetes. 2020. PMID: 32738022
-
Commercial hybrid closed-loop systems available for a patient with type 1 diabetes in 2022.Pediatr Endocrinol Diabetes Metab. 2023;29(1):30-36. doi: 10.5114/pedm.2023.126359. Pediatr Endocrinol Diabetes Metab. 2023. PMID: 37218723 Free PMC article. Review.
Cited by
-
A Response to: Letter to the Editor with Regard to the Cost-Effectiveness of an Advanced Hybrid Closed-Loop System in People with Type 1 Diabetes: A Health Economic Analysis in Sweden.Diabetes Ther. 2022 May;13(5):1125-1129. doi: 10.1007/s13300-022-01252-w. Epub 2022 Apr 11. Diabetes Ther. 2022. PMID: 35403952 Free PMC article. No abstract available.
-
The Influence of Digital Affinity on the Continuous Glucose Monitoring System Choice by People With Type 1 Diabetes.J Diabetes Sci Technol. 2024 Mar;18(2):438-444. doi: 10.1177/19322968221113838. Epub 2022 Jul 20. J Diabetes Sci Technol. 2024. PMID: 35856406 Free PMC article.
-
Prospective Open-Label, Single-Arm, Single-Center Follow-Up Study of the Application of the Advanced Hybrid Closed Loop System in Well-Controlled Children and Adolescents with Type 1 Diabetes.Diabetes Technol Ther. 2022 Nov;24(11):824-831. doi: 10.1089/dia.2022.0148. Epub 2022 Aug 4. Diabetes Technol Ther. 2022. PMID: 35852811 Free PMC article.
-
[Diabetes mellitus in childhood and adolescence (Update 2023)].Wien Klin Wochenschr. 2023 Jan;135(Suppl 1):106-114. doi: 10.1007/s00508-023-02169-5. Epub 2023 Apr 20. Wien Klin Wochenschr. 2023. PMID: 37101031 Free PMC article. German.
-
Hybrid closed-loop systems for managing blood glucose levels in type 1 diabetes: a systematic review and economic modelling.Health Technol Assess. 2024 Dec;28(80):1-190. doi: 10.3310/JYPL3536. Health Technol Assess. 2024. PMID: 39673446 Free PMC article.
Publication types
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous