Extracting Parasite Effects of Electrical Bioimpedance Measurements
- PMID: 33584927
- PMCID: PMC7852010
- DOI: 10.2478/joeb-2018-0016
Extracting Parasite Effects of Electrical Bioimpedance Measurements
Abstract
The objective of this work is to develop a technique for filtering parasitic effects from the impedance spectra (IS) measured in biological material phantoms. IS data are contaminated with unexpected capacitive and inductive effects from cable, input/output amplifiers capacitances, electrode polarization, temperature and contact pressure when collecting data. It is proposed a model which contains an RLC-network in series with the Cole model (RSC), then called RLC-Cole. It was built four circuits composed by resistors, capacitors and inductors. An impedance analyzer (HF2IS) was used to perform the measurements in the frequency range of 1 to 3000 kHz. Data were fitted into the model and comparisons to the nominal values were made. In order to validate the proposed model, a gelatin phantom and a chicken breast muscle impedance spectra were also collected and analyzed. After filtering, Cole fitting was performed. Results showed a maximum root-mean-square error of 1% for the circuits, 2.63% for the gelatin phantom, whereas 2.01% for the chicken breast. The RLC-Cole model could significantly remove parasitic effects out of a tissue impedance spectrum measured by a 4-point electrode probe. This may be highly important in EIS systems whose objective is to discriminate a normal tissue from a cancerous one.
Keywords: Bioimpedance; Cole model; filtering algorithm; parasitic impedance.
© 2018 D. Dutra, P. Bertemes-Filho published by Sciendo.
Conflict of interest statement
Conflict of Interest The authors declare that they have no conflict of interest.
Figures


References
-
- Elwakil A. S., Maundy B.. "Extracting the Cole-Cole impedance model parameters without direct impedance measurement,". Electron. Lett. 2010;46(20):1367. doi: 10.1049/el.2010.1924. - DOI
LinkOut - more resources
Full Text Sources