. 2021 Jun;36(6):2080-2086.
doi: 10.1016/j.arth.2021.01.027. Epub 2021 Jan 23.

Safe Upper Limits of Serum Cobalt and Chromium Levels for a Metal-on-Metal Total Hip Bearing: A 10-Year Follow-Up Study

Affiliations

Safe Upper Limits of Serum Cobalt and Chromium Levels for a Metal-on-Metal Total Hip Bearing: A 10-Year Follow-Up Study

Maarten C Koper et al. J Arthroplasty. 2021 Jun.

Abstract

Background: Long-term survival of metal-on-metal (MoM) prostheses and the development of adverse reaction to metal debris (ARMD) around these bearings are still unclear. Serum levels of cobalt (Co) and chromium (Cr) are used as a screening tool to anticipate failure in MoM bearings and detect ARMD.

Methods: One hundred sixty primary large head MoM prostheses were followed up for 10 years. To estimate the revision risk, the cumulative incidence function (CIF) was used. Subdistribution hazard modeling was used to investigate the associations between cumulative incidence of revision for ARMD and Co levels, Cr levels, gender, age, head size, and cup inclination. Furthermore, the safe upper limits (SULs) for Co and Cr were determined.

Results: Univariate analyses showed an increased risk in revision for ARMD in females (subdistribution hazard ratio [sdHR] 3.43, 95% confidence interval [CI] 1.01-11.7, P = .049) and cup inclination angles over 45° (sdHR 4.70, 95% CI 1.63-13.58, P = .004). In addition, a higher last measured Co level (sdHR 1.05, 95% CI 1.03-1.07, P < .001) and last measured Cr level (sdHR 1.21, 95% CI 1.14-1.29, P < .001) were associated with a higher probability of revision for ARMD. We determined our bearing-specific SULs at 4.1 parts per billion (ppb) and 4.2 ppb for Co and Cr, respectively.

Conclusion: Guidelines regarding follow-up and surveillance should include a complete clinical assessment with bearing-specific SULs of serum metal ion levels. For the M2a-Magnum MoM bearing we advise an SUL for Co and Cr levels of 4.1 and 4.2 ppb, respectively.

Keywords: adverse reaction to metal debris; hip arthroplasty; hip revision; metal-on-metal; safe upper limit.

PubMed Disclaimer