A Comparative Evaluation of Efficacy of Electronic Apex Locator, Digital Radiography, and Conventional Radiographic Method for Root Canal Working Length Determination in Primary Teeth: An In Vitro Study
- PMID: 33623342
- PMCID: PMC7887167
- DOI: 10.5005/jp-journals-10005-1821
A Comparative Evaluation of Efficacy of Electronic Apex Locator, Digital Radiography, and Conventional Radiographic Method for Root Canal Working Length Determination in Primary Teeth: An In Vitro Study
Abstract
Aim: Pulpectomy in the primary tooth has unique challenges due to morphological variations in root pattern and physiological root resorption. The electronic apex locator (EAL) is one of the recent methods to determine the estimation of working length without much radiation exposure to the patient as well as the operator. The present study was undertaken for evaluating and comparing the efficacy of EAL, conventional radiography, digital radiography, and actual visual method for the estimation of in the root canal working length (RCL) in extracted primary teeth.
Materials and methods: Ninety extracted, single-rooted primary teeth were selected. Working length estimation was done with an EAL, conventional, and digital radiographic method, and compared it with an actual visual method.
Results: Accuracy of EAL was observed to be 99.7% followed by digital radiograph (98.1%) and conventional radiograph (96.1%). Both EAL and digital radiographic methods showed a high correlation as compared to conventional. The comparative efficacy of an EAL with a visual method was found to be statistically non-significant (p > 0.005).
Conclusion: Root canal working length determined through the electronic method was found to be an accurate and effective tool in single-rooted primary teeth and can be indicated for clinical implementation in endodontic treatment of primary teeth.
How to cite this article: Sahni A, Kapoor R, Gandhi K, et al. A Comparative Evaluation of Efficacy of Electronic Apex Locator, Digital Radiography, and Conventional Radiographic Method for Root Canal Working Length Determination in Primary Teeth: An In Vitro Study. Int J Clin Pediatr Dent 2020;13(5):523-528.
Keywords: Digital radiograph; Electronic apex locator; Primary tooth; Working length.
Copyright © 2020; Jaypee Brothers Medical Publishers (P) Ltd.
Conflict of interest statement
Source of support: Nil Conflict of interest: None
Figures







Similar articles
-
Comparative Evaluation of the Accuracy of Electronic Apex Locator and Digital Radiography for Working Length Determination in Primary Teeth: A Systematic Review.J Dent (Shiraz). 2024 Sep 1;25(3):203-214. doi: 10.30476/dentjods.2023.97323.2006. eCollection 2024 Sep. J Dent (Shiraz). 2024. PMID: 39371951 Free PMC article. Review.
-
Comparison of Radiovisiography, an Apex Locator and an Integrated Endomotor-inbuilt Apex Locator in Primary Teeth Endometrics.Int J Clin Pediatr Dent. 2022;15(Suppl 1):S18-S21. doi: 10.5005/jp-journals-10005-2123. Int J Clin Pediatr Dent. 2022. PMID: 35645506 Free PMC article.
-
Evaluation of the Working Length Determination Accuracy by Cone-beam Computed Tomography in Primary Teeth.Int J Clin Pediatr Dent. 2022;15(Suppl 1):S92-S96. doi: 10.5005/jp-journals-10005-2330. Int J Clin Pediatr Dent. 2022. PMID: 35645504 Free PMC article.
-
A Comparative Evaluation of Accuracy of New-generation Electronic Apex Locator with Conventional Radiography to determine Working Length in Primary Teeth: An in vivo Study.Int J Clin Pediatr Dent. 2017 Jan-Mar;10(1):34-36. doi: 10.5005/jp-journals-10005-1403. Epub 2017 Feb 27. Int J Clin Pediatr Dent. 2017. PMID: 28377652 Free PMC article.
-
Determination of working length for teeth with wide or immature apices: a review.Int Endod J. 2013 Jun;46(6):483-91. doi: 10.1111/iej.12032. Epub 2012 Dec 7. Int Endod J. 2013. PMID: 23216284 Review.
Cited by
-
The evaluation of the accuracy of a wireless electronic apex locator in primary molar teeth.BMC Oral Health. 2024 Dec 31;24(1):1580. doi: 10.1186/s12903-024-05387-z. BMC Oral Health. 2024. PMID: 39741258 Free PMC article.
-
Knowledge, awareness, and practice survey on conventional radiographic methods and processing among dental students.J Adv Pharm Technol Res. 2022 Nov;13(Suppl 1):S6-S11. doi: 10.4103/japtr.japtr_299_22. Epub 2022 Nov 30. J Adv Pharm Technol Res. 2022. PMID: 36643141 Free PMC article.
-
Comparative Evaluation of the Accuracy of Electronic Apex Locator and Digital Radiography for Working Length Determination in Primary Teeth: A Systematic Review.J Dent (Shiraz). 2024 Sep 1;25(3):203-214. doi: 10.30476/dentjods.2023.97323.2006. eCollection 2024 Sep. J Dent (Shiraz). 2024. PMID: 39371951 Free PMC article. Review.
-
How Do Different Image Modules Impact the Accuracy of Working Length Measurements in Digital Periapical Radiography? An In Vitro Study.Diagnostics (Basel). 2025 Jan 28;15(3):305. doi: 10.3390/diagnostics15030305. Diagnostics (Basel). 2025. PMID: 39941235 Free PMC article.
References
-
- Ingle J, Backland LK. Pediatric Endodontics. Endodontics. 5th ed., Philadelphia: Lea and Febiger; 2002. p. 861.
-
- Katz A, Mass E, Kaufman AY. Electronic apex locator: a useful tool for root canal treatment in the primary dentition. ASDC J Dent Child. 1996;63:414–417. - PubMed
LinkOut - more resources
Full Text Sources
Miscellaneous