Perinatal outcome after vacuum assisted delivery with digital feedback on traction force; a randomised controlled study
- PMID: 33637058
- PMCID: PMC7913459
- DOI: 10.1186/s12884-021-03604-z
Perinatal outcome after vacuum assisted delivery with digital feedback on traction force; a randomised controlled study
Abstract
Background: Low and mid station vacuum assisted deliveries (VAD) are delicate manual procedures that entail a high degree of subjectivity from the operator and are associated with adverse neonatal outcome. Little has been done to improve the procedure, including the technical development, traction force and the possibility of objective documentation. We aimed to explore if a digital handle with instant haptic feedback on traction force would reduce the neonatal risk during low or mid station VAD.
Methods: A two centre, randomised superiority trial at Karolinska University Hospital, Sweden, 2016-2018. Cases were randomised bedside to either a conventional or a digital handle attached to a Bird metal cup (50 mm, 80 kPa). The digital handle measured applied force including an instant notification by vibration when high levels of traction force were predicted according to a predefined algorithm. Primary outcome was a composite of hypoxic ischaemic encephalopathy, intracranial haemorrhage, seizures, death and/or subgaleal hematoma. Three hundred eighty low and mid VAD in each group were estimated to decrease primary outcome from six to 2 %.
Results: After 2 years, an interim analyse was undertaken. Meeting the inclusion criteria, 567 vacuum extractions were randomized to the use of a digital handle (n = 296) or a conventional handle (n = 271). Primary outcome did not differ between the two groups: (2.7% digital handle vs 2.6% conventional handle). The incidence of primary outcome differed significantly between the two delivery wards (4% vs 0.9%, p < 0.05). A recalculation of power revealed that 800 cases would be needed in each group to show a decrease in primary outcome from three to 1 %. This was not feasible, and the study therefore closed.
Conclusions: The incidence of primary outcome was lower than estimated and the study was underpowered. However, the difference between the two delivery wards might reflect varying degree of experience of the technical equipment. An objective documentation of the extraction procedure is an attractive alternative in respect to safety and clinical training. To demonstrate improved safety, a multicentre study is required to reach an adequate cohort. This was beyond the scope of the study.
Trial registration: ClinicalTrials.gov NCT03071783 , March 1, 2017, retrospectively registered.
Keywords: Haptic feedback; Neonatal outcome; Traction force; Vacuum assisted delivery.
Conflict of interest statement
GA, MW and KY are owners of Genit Innovation, which is developing the technical device described in this work. KP and SR have no competing interests to declare. All authors have completed the ICMJE uniform disclosure.
Figures
Similar articles
-
Traction force profile in children with severe perinatal outcomes delivered with a digital vacuum extraction handle: A case-control study.Acta Obstet Gynecol Scand. 2022 Nov;101(11):1238-1244. doi: 10.1111/aogs.14444. Epub 2022 Aug 28. Acta Obstet Gynecol Scand. 2022. PMID: 36030477 Free PMC article.
-
Association of traction force and adverse neonatal outcome in vacuum-assisted vaginal delivery: A prospective cohort study.Acta Obstet Gynecol Scand. 2020 Dec;99(12):1710-1716. doi: 10.1111/aogs.13952. Epub 2020 Sep 9. Acta Obstet Gynecol Scand. 2020. PMID: 32644188
-
Traction force during vacuum extraction: a prospective observational study.BJOG. 2015 Dec;122(13):1809-16. doi: 10.1111/1471-0528.13222. Epub 2015 Jan 5. BJOG. 2015. PMID: 25558833
-
Neonatal subgaleal hemorrhage and its relationship to delivery by vacuum extraction.Obstet Gynecol Surv. 2003 Oct;58(10):687-93. doi: 10.1097/01.OGX.0000086420.13848.89. Obstet Gynecol Surv. 2003. PMID: 14515066 Review.
-
[Neonatal consequences of vacuum-assisted delivery].Ugeskr Laeger. 2010 Aug 30;172(35):2376-81. Ugeskr Laeger. 2010. PMID: 20825742 Review. Danish.
Cited by
-
Can We Predict Feto-Maternal Adverse Outcomes of Vacuum Extraction?Geburtshilfe Frauenheilkd. 2022 Sep 7;82(11):1274-1282. doi: 10.1055/a-1904-6025. eCollection 2022 Nov. Geburtshilfe Frauenheilkd. 2022. PMID: 36339635 Free PMC article.
-
Instruments for assisted vaginal birth.Cochrane Database Syst Rev. 2021 Sep 24;9(9):CD005455. doi: 10.1002/14651858.CD005455.pub3. Cochrane Database Syst Rev. 2021. PMID: 34559884 Free PMC article.
-
Traction force profile in children with severe perinatal outcomes delivered with a digital vacuum extraction handle: A case-control study.Acta Obstet Gynecol Scand. 2022 Nov;101(11):1238-1244. doi: 10.1111/aogs.14444. Epub 2022 Aug 28. Acta Obstet Gynecol Scand. 2022. PMID: 36030477 Free PMC article.
References
-
- Malmstrom T. Vacuum extractor, an obstetrical instrument. Acta Obstet Gynecol Scand Suppl. 1954;33(4):1–31. - PubMed
-
- The Swedish National Board of Health. Welfare T. Statistikdatabas för graviditeter, förlossningar och nyfödda. 2016.
-
- Vacca A. Reducing the risks of a vacuum delivery. Fetal Med Rev. 2006;17(4):301–315. doi: 10.1017/S0965539506001823. - DOI
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical