Early outcomes of Jotec inner-branched endografts in complex endovascular aortic aneurysm repair
- PMID: 33647435
- DOI: 10.1016/j.jvs.2021.01.067
Early outcomes of Jotec inner-branched endografts in complex endovascular aortic aneurysm repair
Abstract
Objective: Complex aortic endografts have evolved to include inner branches to overcome specific challenges with existing technologies. We have reported the early outcomes of endovascular aortic aneurysm repair (EVAR) using a Jotec inner branched endograft (iBEVAR).
Methods: All patients who had undergone complex EVARs using extra-design engineering iBEVAR (Jotec GmbH, Hechingen, Germany) from 2018 to 2020 at a single center were reviewed. The patient demographics, cardiovascular risk factors, anatomic features of the aneurysms, and target vessels were recorded. The reasons for using inner branches instead of fenestrated and standard branched endografts and the procedural details, outcomes, and reintervention during follow-up were examined.
Results: A total of 110 patients were treated with branched and fenestrated endografts during the study period, of whom 18 patients had had a patient-specific custom-made iBEVAR endograft with downward inner branches. The technical success rate was 100%. A total of 68 target vessels were cannulated, and bridging stent-grafts were placed successfully in all. The reasons for choosing the iBEVAR design included unfavorable target vessel trajectory for fenestrated repair (n = 15), excessive infrarenal aortic angulation and/or adverse iliac access vessels for fenestrated repair (n = 11), the presence of a narrow aortic lumen (n = 14), and/or to reduce aortic coverage compared with that with standard outer branched repair (n = 14). We also used iBEVAR to treat type Ia endoleaks after failed EVAR with a short main body (n = 5). The median contrast volume used was 120 mL (range, 48-200 mL), with a median fluoroscopy screening time of 66 minutes (range, 35-136 minutes) and a median dose-area product of 17,832 dGy∙cm2 (range, 8260-55,070 dGycm2). No 30-day mortality and no major complications occurred. One early intervention was required for a suspected type Ib endoleak from an iliac limb and one late intervention for in-stent stenosis in a renal bridging stent-graft. One patient had died of non-aortic-related causes at 3 months. All other patients continued with follow-up with their aneurysms excluded, patent target vessels, and no type I or III endoleak identified at a median follow-up of 12 months (range, 1-26 months).
Conclusions: The use of Jotec extra-design engineering endografts incorporating downward inner branches resulted in satisfactory early outcomes with a low reintervention rate. The technology has the potential to be a useful addition to our armamentarium for treating complex aortic endografts; however, long-term outcomes data are needed.
Keywords: Aortic endografts; Branched EVAR; Fenestrated EVAR; Inner branched EVAR; Revision EVAR.
Copyright © 2021 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.
Comment in
-
Innovative approaches and new endograft designs to perform complex endovascular thoracoabdominal aortic aneurysm repair.J Vasc Surg. 2021 Sep;74(3):880-881. doi: 10.1016/j.jvs.2021.02.018. J Vasc Surg. 2021. PMID: 34425955 No abstract available.
Similar articles
-
Early and Mid-Term Outcomes of the Inverted Limb Configuration Below Fenestrated and Branched Endografts: Experience from Two European Centers.J Endovasc Ther. 2024 Jun;31(3):410-420. doi: 10.1177/15266028221125158. Epub 2022 Oct 2. J Endovasc Ther. 2024. PMID: 36189939
-
Two-year evaluation of fenestrated and parallel branch endografts for the treatment of juxtarenal, suprarenal, and thoracoabdominal aneurysms at a single institution.J Vasc Surg. 2020 Jan;71(1):15-22. doi: 10.1016/j.jvs.2019.03.058. Epub 2019 Nov 9. J Vasc Surg. 2020. PMID: 31718954
-
Effect of aortic angulation on the outcomes of fenestrated-branched endovascular aortic repair.J Vasc Surg. 2021 Aug;74(2):372-382.e3. doi: 10.1016/j.jvs.2021.01.027. Epub 2021 Feb 4. J Vasc Surg. 2021. PMID: 33548434
-
Endovascular treatment of complex abdominal and thoracoabdominal type IV aortic aneurysms with fenestrated technology.J Cardiovasc Surg (Torino). 2017 Aug;58(4):574-590. doi: 10.23736/S0021-9509.16.09098-4. Epub 2016 Feb 4. J Cardiovasc Surg (Torino). 2017. PMID: 26845537 Review.
-
A Comprehensive Review of In Situ Fenestration of Aortic Endografts.Eur J Vasc Endovasc Surg. 2016 Dec;52(6):787-800. doi: 10.1016/j.ejvs.2016.10.001. Epub 2016 Nov 11. Eur J Vasc Endovasc Surg. 2016. PMID: 27843111
Cited by
-
Single-center initial experience with inner-branch complex EVAR in 44 patients.Front Cardiovasc Med. 2023 Jun 15;10:1188501. doi: 10.3389/fcvm.2023.1188501. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 37396572 Free PMC article.
-
Early Experience with Inner Branch Stent-Graft System for Endovascular Repair of Thoraco-Abdominal and Pararenal Abdominal Aortic Aneurysm.Diagnostics (Basel). 2024 Nov 21;14(23):2612. doi: 10.3390/diagnostics14232612. Diagnostics (Basel). 2024. PMID: 39682521 Free PMC article.
-
Treatment of late-presenting malperfusion and aneurysmal degeneration in chronic type B aortic dissection with an off-the-shelf four-vessel inner-branched endograft.J Vasc Surg Cases Innov Tech. 2023 Jul 23;9(3):101273. doi: 10.1016/j.jvscit.2023.101273. eCollection 2023 Sep. J Vasc Surg Cases Innov Tech. 2023. PMID: 37635743 Free PMC article.
-
Staged Hybrid Treatment with Branched Endovascular Aneurysm Repair of a Thoracoabdominal Aortic Aneurysm in the Presence of a Total Infrarenal Aortoiliac Occlusion.Vasc Specialist Int. 2021 Dec 31;37:43. doi: 10.5758/vsi.210064. Vasc Specialist Int. 2021. PMID: 34983027 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources