Immunoturbidimetric assays for serum apolipoproteins A1 and B using Cobas Bio centrifugal analyser
- PMID: 3366936
- PMCID: PMC1141480
- DOI: 10.1136/jcp.41.4.471
Immunoturbidimetric assays for serum apolipoproteins A1 and B using Cobas Bio centrifugal analyser
Abstract
Immunoturbidimetric assays for measuring the apolipoproteins A1 and B using the Cobas Bio centrifugal analyser are described. The methods were specific, offered good sensitivity (less than 0.05 g/l) and intrabatch variability, with coefficients of variation between 2.4% and 3.5%, and were cost effective. Reference ranges were calculated for a group of civil servants, aged 35 to 55 years.
References
MeSH terms