Effect of a pocket compression device on hematomas, skin reactions, and comfort in patients receiving a cardiovascular implantable electronic device: a randomized controlled trial
- PMID: 33675448
- DOI: 10.1007/s10840-021-00973-5
Effect of a pocket compression device on hematomas, skin reactions, and comfort in patients receiving a cardiovascular implantable electronic device: a randomized controlled trial
Abstract
Aim/objective: Aim to evaluate the safety and effectiveness of a pocket compression device in patients undergoing cardiovascular implantable electronic device (CIED) implantation.
Material and methods: This randomized controlled trial included 54 patients in the control group (elastic adhesive tape) and 54 patients in the experimental group (pocket compression device). A sandbag was positioned over the pocket after CIED implantation, using elastic adhesive tape in the control group and a novel pocket compression device in the experimental group. The primary outcome was the incidence of pocket hematomas, and the secondary outcomes were adverse skin reactions, the need for sandbag positional adjustments, and patient comfort.
Results: The distributions of age, sex, and clinical and procedure characteristics was not significantly different between the two groups (P>0.05). The occurrence of hematomas was lower when using the pocket compression device than in the control group (13.0% vs. 44.4%, P<0.001). The incidence rates of hematomas of grades 1 and the grade 2 hematomas were significantly lower in the experimental group than in the control group (13.0% vs. 31.5%, 0% vs. 11.1% respectively, [all P<0.05]). The incidence rates of adverse skin reactions and sandbag positional adjustments were much lower and the patient comfort score was significantly higher in the experimental group than in the control group (all P<0.001).
Conclusion: A pocket compression device can significantly decrease the incidence of pocket hematomas, adverse skin reactions, and positional adjustments when patients are undergoing CIED, which can improve patient comfort and decrease the nursing workload.
Clinicaltrials: GOV : NCT04389398.
Keywords: and comfort; cardiovascular implantable electronic device; hematomas; pocket compression device; skin reactions.
© 2021. Springer Science+Business Media, LLC, part of Springer Nature.
References
-
- Tracy CM, Epstein AE, Darbar D, DiMarco JP, Dunbar SB, Estes NA 3rd, et al. 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Thorac Cardiovasc Surg. 2012;144(6):e127–45. - DOI
-
- Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, et al. ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J. 2013;34(29):2281–329. - DOI
-
- Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503–51. - DOI
-
- Yang X, Wang Z, Zhang Y, Yin X, Hou Y. The safety and efficacy of antithrombotic therapy in patients undergoing cardiac rhythm device implantation: a meta-analysis. Europace. 2015;17(7):1076–84. - DOI
-
- Essebag V, Verma A, Healey JS, Krahn AD, Kalfon E, Coutu B, et al. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION study. J Am Coll Cardiol. 2016;67(11):1300–8. - DOI