Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Feb 15;6(8):5805-5811.
doi: 10.1021/acsomega.0c06248. eCollection 2021 Mar 2.

Effect of 18-Crown-6 on Oxytocin Stability in Aqueous Buffer Solutions

Affiliations

Effect of 18-Crown-6 on Oxytocin Stability in Aqueous Buffer Solutions

Mostafa Ghasemisarabbadieh et al. ACS Omega. .

Abstract

In this study, the effect of 18-crown-6 on the stability of oxytocin in aqueous solution was explored. The study found that while 12-crown-4 and 15-crown-5 do not stabilize oxytocin, 18-crown-6 does have a stabilizing effect in citrate/phosphate buffer at pH 4.5. However, in acetate buffer at the same pH, the presence of 18-crown-6 had a destabilizing effect, possibly leading to a different degradation pathway. Both the stabilizing and destabilizing effects, depending on the buffer used, are concentration dependent where a higher concentration of 18-crown-6 is linked to a stronger effect. It is hypothesized that this effect may be linked to 18-crown-6 binding to the protonated ammonium group of oxytocin. Upon changing the mobile phase used in high-performance liquid chromatography experiments, we observed evidence supporting this binding hypothesis. When an acidic mobile phase was used (0.01% trifluoroacetic acid (TFA)), a partial shift in oxytocin retention time was observed for samples in acetate buffers in the presence of 18-crown-6 when using a 150 mm column (C18). The amount of the peak that shifted depended on the 18-crown-6 concentration used. A similar shift in oxytocin peak retention time was observed for samples in both acetate and citrate/phosphate buffers when using a 250 mm column (C18), but the peak completely shifted in those samples. When using an even more acidic mobile phase (0.1% TFA), the oxytocin peaks all had the same retention time again. Ultraviolet and nuclear magnetic resonance spectroscopy experiments also showed that the presence of 18-crown-6 has an observable effect on the resulting oxytocin spectra.

PubMed Disclaimer

Conflict of interest statement

The authors declare no competing financial interest.

Figures

Figure 1
Figure 1
Hypothesized stabilizing interaction between 18-crown-6 (blue) and the protonated ammonium group (red) on the oxytocin molecule.
Figure 2
Figure 2
HPLC traces of oxytocin after 14 days of storage at 50 °C in the presence and absence of 2% w/v of 18-crown-6 in both acetate and citrate/phosphate buffer.
Figure 3
Figure 3
Ultraviolet (UV) trace from HPLC runs of an oxytocin standard (black), an oxytocin sample with 0.1% w/v of 18-crown-6 added (blue), and an oxytocin sample with 1.0% w/v of 18-crown-6 added (red). These samples were prepared in acetate buffer.
Figure 4
Figure 4
HPLC traces of samples showing an oxytocin peak shift when in the presence of 18-crown-6. The samples were prepared in citrate/phosphate buffer (left) and in acetate buffer (right). The mobile phase used was A: 0.01% TFA in H2O and B: 0.01% TFA in 70% MeCN: 30% H2O.
Figure 5
Figure 5
HPLC traces of samples showing no oxytocin peak shift regardless of the presence of 18-crown-6. The samples were prepared in citrate/phosphate buffer (left) and acetate buffer (right). The mobile phase used was A: 0.1% TFA in H2O and B: 0.1% TFA in 70% MeCN: 30% H2O.
Figure 6
Figure 6
UV absorption spectra for oxytocin, 18-crown-6, a solution with both oxytocin and 18-crown-6, and a solution with oxytocin, 18-crown-6, and potassium chloride.
Figure 7
Figure 7
Stacked NMR spectra of pure oxytocin (blue) and a mixture of oxytocin and 18-crown-6 (black) in D2O.

Similar articles

References

    1. Vogel J. P.; Williams M.; Gallos I.; Althabe F.; Oladapo O. T.; WHO Recommendations Uterotonics for the Prevention of Postpartum Haemorrhage; Geneva, 2018. - PMC - PubMed
    1. Say L.; Chou D.; Gemmill A.; Tunçalp Ö.; Moller A.; Daniels J.; Gülmezoglu A. M.; Temmerman M.; Alkema L. Global Causes of Maternal Death: A WHO Systematic Analysis. Lancet Glob. Health 2014, 2, e323–e333. 10.1016/S2214-109X(14)70227-X. - DOI - PubMed
    1. Winkler M.; Rath W. A Risk-Benefit Assessment of Oxytocics in Obstetric Practice. Drug Saf. 1999, 20, 323–345. 10.2165/00002018-199920040-00003. - DOI - PubMed
    1. Van Dongen P. W. J.; Van Roosmalen J.; De Boer C. N.; Van Rooij J. Oxytocics for the Prevention of Post-Partum Haemorrhages. Pharm. Weekbl. Sci. Ed 1991, 13, 238–243. - PubMed
    1. Olff M.; Frijling J. L.; Kubzansky L. D.; Bradley B.; Ellenbogen M. A.; Cardoso C.; Bartz J. A.; Yee J. R.; van Zuiden M. The Role of Oxytocin in Social Bonding, Stress Regulation and Mental Health: An Update on the Moderating Effects of Context and Interindividual Differences. Psychoneuroendocrinol. 2013, 38, 1883–1894. 10.1016/j.psyneuen.2013.06.019. - DOI - PubMed

LinkOut - more resources