A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit
- PMID: 33683733
- DOI: 10.1111/iej.13507
A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit
Abstract
Aim: To evaluate several factors that might interfere with the use of electronic root canal length measurement devices (ERCLMDs) in a laboratory setting, including two different embedding media (alginate and electroconductive gel), three different types of devices and the radiographic view on the assessment of the electronic readings.
Methodology: Thirty single-rooted extracted human mandibular premolars were selected. After access and canal pre-flaring, a size 10 K-file was inserted in the canal up to the major apical foramen under magnification (×10), and this length was recorded as the actual length (AL) of the canal. Teeth were mounted on a specific endodontic training kit platform (ProTrain). Two different embedding media (electroconductive gel and alginate) and three different ERCLMDs: Apex ID (SybronEndo, Glendora, CA, USA), CanalPro Apex Locator (Coltene-Endo, Cuyahoga Falls, OH, USA) and the Root ZX II (J. Morita, Tokyo, Japan), were tested. Following the measurement at the 0.5 mark of each ERCLMD using alginate, two periapical radiographs (bucco-lingual and proximal views) were taken. The difference between the electronic readings and the AL, as well as the distance from the file tip to the apex of the roots on the radiographs, was calculated. Data were analysed statistically (two-way anova, Tukey's and chi-squared tests) at 5% significance level.
Results: The incidence of over-extended readings was significantly greater (P < 0.05) when using the ProTrain electroconductive gel, except for Root ZX II at the 0.5 mark. Root ZX, CanalPro and Apex ID had similar accuracy when the 0.0/APEX mark was used to locate the foramen. For the radiographic method, no difference was found between the devices or radiographic views.
Conclusion: Compared with alginate, the conductive gel of the ProTrain kit negatively affected most of the electronic readings. The three evaluated ERCLMDs had similar accuracy in locating the apical foramen when using the mark 0.0/APEX and alginate as the embedding media. Although the ProTrain platform allows radiographs to be exposed in both bucco-lingual and proximal views, no difference was found between the views when evaluating the apical limit provided by the 0.5 mark in mandibular premolars embedded in alginate.
Keywords: electronic root canal length measurement device; embedding media; endodontics; tooth apex; working length.
© 2021 International Endodontic Journal. Published by John Wiley & Sons Ltd.
References
-
- Aurelio JA, Nahmias Y, Gerstein H (1983) A model for demonstrating an electronic canal length measuring device. Journal of Endodontics 9, 568-9.
-
- Baldi JV, Victorino FR, Bernardes RA et al. (2007) Influence of embedding media on the assessment of electronic apex locators. Journal of Endodontics 33, 476-9.
-
- Chen E, Kaing S, Mohan H, Ting SY, Wu J, Parashos P (2011) An ex vivo comparison of electronic apex locator teaching models. Journal of Endodontics 37, 1147-51.
-
- Cianconi L, Angotti V, Felici R, Conte G, Mancini M (2010) Accuracy of three electronic apex locators compared with digital radiography: an ex vivo study. Journal of Endodontics 36, 2003-7.
-
- Connert T, Judenhofer MS, Hülber JM et al. (2018) Evaluation of the accuracy of nine electronic apex locators by using Micro-CT. International Endodontic Journal 51, 223-32.