Management of Nellix migration and type Ia endoleak from proximal endovascular aneurysm sealing relining to late open conversion
- PMID: 33684472
- DOI: 10.1016/j.jvs.2021.02.035
Management of Nellix migration and type Ia endoleak from proximal endovascular aneurysm sealing relining to late open conversion
Abstract
Background: Despite promising early results, midterm failures of the Nellix endovascular aneurysm sealing (EVAS) system (Endologix Inc, Irvine, Calif) have been reported at higher than expected rates. The management of proximal endoleaks and migration differs from those after conventional endovascular aortic aneurysm repair (EVAR) owing to the peculiar design of the Nellix device. In the present study, we report a monocentric experience in the management of EVAS complications using various techniques. We also performed a comprehensive review of the relevant literature on both open surgical and endovascular management of proximal failure of EVAS from the MEDLINE database.
Methods: We retrospectively analyzed the reinterventions for type Ia endoleak and migration after elective infrarenal EVAS at our institution. We collected preoperative, intraoperative, and follow-up data. Open and endovascular techniques are described. Overall survival, aortic-related mortality, and the technical success rate (rate of exclusion of endoleaks) with endovascular techniques were the primary outcomes.
Results: We performed 101 infrarenal elective EVAS procedures from 2013 to 2018. Of the 101 patients, 20 (19.8%) had required reintervention for proximal sealing failure. The indications were type Ia (Is2, Is3) endoleak, migration >5 mm, sac expansion >5 mm, and secondary rupture. Of the 20 patients, 6 (30%) were treated with endovascular techniques-2 with a chimney Nellix-in-Nellix application and 4 with proximal relining with a covered stent. The remaining 14 patients (70%) were treated with late open conversion (OC). The average time from EVAS to reintervention was 36.1 months (range, 3-65 months). Six patients (30%) had undergone OC in an emergent setting because of secondary rupture. The technical success rate for the patients treated with endovascular reinterventions was 100%. The 30-day mortality was 20% (4 of 20), all emergent cases (four of six emergent repairs; 67%). The overall survival for the 20 patients was 75% (n = 15) at a mean follow-up of 15.1 months (range, 2-47 months). One patient had died after 7 months of non-aortic-related causes.
Conclusions: The high reintervention rate of the Nellix graft mandates careful evaluation for its further use with the revised instructions for use, and it should not be used off-label. OC remains the strategy of choice when managing Nellix proximal sealing failures in fit patients. Chimney Nellix-in-Nellix application and transcatheter embolization are feasible alternative techniques. Proximal relining also appears to be an effective alternative to more complex interventions, although it requires further studies for validation.
Keywords: Abdominal aortic aneurysm; EVAS; Endoleak; Graft failure; Migration; Nellix.
Copyright © 2021 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Long-term failure after endovascular aneurysm sealing in a real-life, single-center experience with the Nellix endograft.J Vasc Surg. 2021 Jun;73(6):1958-1965.e1. doi: 10.1016/j.jvs.2020.11.029. Epub 2020 Dec 2. J Vasc Surg. 2021. PMID: 33278539
-
Preliminary outcome of Nellix-in-Nellix extensions in patients treated with failed endovascular aneurysm sealing.J Vasc Surg. 2019 Oct;70(4):1099-1106. doi: 10.1016/j.jvs.2019.01.044. Epub 2019 Mar 29. J Vasc Surg. 2019. PMID: 30935765
-
Preliminary results of endovascular aneurysm sealing from the multicenter Italian Research on Nellix Endoprosthesis (IRENE) study.J Vasc Surg. 2018 May;67(5):1397-1403. doi: 10.1016/j.jvs.2017.09.032. Epub 2017 Dec 11. J Vasc Surg. 2018. PMID: 29242065
-
Endovascular aneurysm sealing as an alternative for the treatment of failed endovascular aneurysm repair.Vascular. 2020 Jun;28(3):251-258. doi: 10.1177/1708538119897525. Epub 2020 Jan 2. Vascular. 2020. PMID: 31896300
-
Management of Type IA Endoleak After EVAR by Explantation or Custom Made Fenestrated Endovascular Aortic Aneurysm Repair.Eur J Vasc Endovasc Surg. 2021 Apr;61(4):571-578. doi: 10.1016/j.ejvs.2020.10.033. Epub 2021 Jan 4. Eur J Vasc Endovasc Surg. 2021. PMID: 33414067 Review.
Cited by
-
ALTURA™ Stent Graft Shortening and Its Implications After EVAR.J Clin Med. 2025 Feb 11;14(4):1157. doi: 10.3390/jcm14041157. J Clin Med. 2025. PMID: 40004688 Free PMC article.
-
Open Surgical Conversion After Failed Endovascular Aneurysm Sealing.Vasc Health Risk Manag. 2025 Jun 13;21:477-488. doi: 10.2147/VHRM.S513365. eCollection 2025. Vasc Health Risk Manag. 2025. PMID: 40535154 Free PMC article.
-
A Case of Late Type Ia Endoleak After Endovascular Aneurysm Sealing Using the Nellix System: Proximal Extension with Triple Chimney and Gutter Endoleak Embolization.Ann Vasc Dis. 2021 Dec 25;14(4):393-395. doi: 10.3400/avd.cr.21-00092. Ann Vasc Dis. 2021. PMID: 35082948 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources