Bilateral Mechanically-Assisted Crevice Corrosion Resulting in Femoral Stem-Head Dissociation in Metal-on-Polyethylene Total Hip Arthroplasty
- PMID: 33718553
- PMCID: PMC7921711
- DOI: 10.1016/j.artd.2020.12.018
Bilateral Mechanically-Assisted Crevice Corrosion Resulting in Femoral Stem-Head Dissociation in Metal-on-Polyethylene Total Hip Arthroplasty
Abstract
There are emerging reports of complications related to corrosion in modular femoral components. We report a unique case describing an 83-year-old man with bilateral mechanically-assisted crevice corrosion in hip replacements performed 10 years previously, by the same surgeon using the same size 3 Accolade TMZF stem and same 44-mm CoCr LFIT V40 head (Stryker Orthopedics, Mahwah NJ). Our patient presented with complete femoral stem-head complex dissociation of his right hip and elevated serum cobalt and chromium levels. He subsequently underwent right hip revision arthroplasty while his contralateral hip is monitored closely as an outpatient. This case helps to define the poorly understood mechanisms and component design factors implicated in this emerging issue. We also hope to provoke discussions about guidelines for monitoring and revising failing metal-on-polyethylene arthroplasty systems.
Keywords: Accolade TMZF; LFIT V40 cobalt-chrome head; Mechanically-assisted cervice corrosion; Metal on polyethylene arthroplasty; Modular femoral head dissociation.
© 2020 The Authors.
Figures



Similar articles
-
Catastrophic failure of the Accolade I hip arthroplasty stem: a retrieval analysis study.Hip Int. 2020 Jul;30(4):481-487. doi: 10.1177/1120700020919665. Epub 2020 Apr 23. Hip Int. 2020. PMID: 32323587
-
Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.J Arthroplasty. 2017 Sep;32(9):2857-2863. doi: 10.1016/j.arth.2017.03.075. Epub 2017 Apr 8. J Arthroplasty. 2017. PMID: 28478184 Free PMC article.
-
Head Taper Corrosion Causing Head Bottoming Out and Consecutive Gross Stem Taper Failure in Total Hip Arthroplasty.J Arthroplasty. 2018 Nov;33(11):3581-3590. doi: 10.1016/j.arth.2018.07.017. Epub 2018 Jul 24. J Arthroplasty. 2018. PMID: 30100136
-
Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature.Int Orthop. 2020 Apr;44(4):609-621. doi: 10.1007/s00264-019-04474-z. Epub 2020 Jan 4. Int Orthop. 2020. PMID: 31900575
-
Corrosion at the head-neck interface of current designs of modular femoral components: essential questions and answers relating to corrosion in modular head-neck junctions.Bone Joint J. 2016 May;98-B(5):579-84. doi: 10.1302/0301-620X.98B5.35592. Bone Joint J. 2016. PMID: 27143725 Review.
Cited by
-
Large head metal-on-metal bearing surface with a TMZF titanium alloy femoral stem with high rates of revision and trunnion failure.J Orthop. 2023 Nov 7;46:164-168. doi: 10.1016/j.jor.2023.11.002. eCollection 2023 Dec. J Orthop. 2023. PMID: 38031627 Free PMC article.
-
Unusual Case of a Trunnion Fracture Following a Revision Hip Arthroplasty Surgery: A Brief Review of the Literature and Discussion About Causes of Failure.Arthroplast Today. 2022 Oct 22;18:103-106. doi: 10.1016/j.artd.2022.09.012. eCollection 2022 Dec. Arthroplast Today. 2022. PMID: 36304696 Free PMC article.
References
-
- Srinivasan A., Jung E., Levine B.R. Modularity of the femoral component in total hip arthroplasty. J Am Acad Orthop Surg. 2012;20(4):214. - PubMed
-
- Krishnan H., Krishnan S.P., Blunn G., Skinner J.A., Hart A.J. Modular neck femoral stems. Bone Joint J. 2013;95-b(8):1011. - PubMed
-
- Lucas L.C., Buchanan R.A., Lemons J.E. Investigations on the galvanic corrosion of multialloy total hip prostheses. J Biomed Mater Res. 1981;15(5):731. - PubMed
-
- Whitehouse M.R., Endo M., Zachara S. Adverse local tissue reactions in metal-on-polyethylene total hip arthroplasty due to trunnion corrosion: the risk of misdiagnosis. Bone Joint J. 2015;97-b(8):1024. - PubMed
Publication types
LinkOut - more resources
Full Text Sources
Other Literature Sources