Quantitative assessment of infusion pump-mediated haemolysis in feline packed red blood cell transfusions
- PMID: 33719675
- PMCID: PMC10812153
- DOI: 10.1177/1098612X21999990
Quantitative assessment of infusion pump-mediated haemolysis in feline packed red blood cell transfusions
Abstract
Objectives: Haemolysis caused by the use of peristaltic infusion pumps (PIPs) has been described in human and canine packed red blood cells (pRBCs). The aim of this study was to evaluate the effects of two different linear PIPs on the haemolysis of feline pRBC units stored for a long time.
Methods: Feline pRBC units stored with adenine, dextrose, mannitol and sodium chloride (SAGM) were manufactured. After 35-42 days of storage at 2-4°C, a line administration system with a 180 µm filter was attached to every pRBC bag, the system was drained by gravity alone (8 drops/min) and a 1.3 ml sample was collected (G). A NIKI V4 pump was then used at a flow rate of 25 ml/h, the flow was stopped when the infusion system was filled with blood coming from the infusion pump and another 1.3 ml sample was collected (NK). Finally, an Infusomat FmS pump was evaluated, collecting another 1.3 ml sample (IM). Packed cell volume (PCV) was measured in all samples by microhaematocrit centrifugation, total haemoglobin (HGB) was measured using a specific haemoglobin analyser and, after centrifugation, free HGB was determined by spectrophotometry. The percentage of haemolysis was calculated. Friedman's test was used to compare the samples.
Results: Fifteen feline pRBC units were evaluated. The average degree of haemolysis for sample G (gravity-assisted) was 1.12%. Comparison of the degree of gravity-assisted haemolysis with haemolysis in PIP NK (1.13%) and IM (1.14%) samples revealed no significant differences, with differences of only 0.01% and 0.02%, respectively.
Conclusions and relevance: The results of this study demonstrate that the use of two common PIPs in veterinary hospitals does not produce levels of haemolysis that are significantly different than that caused by gravity alone during transfusion of feline pRBCs at a rate of 25 ml/h.
Keywords: Infusomat FmS; NIKI V4; Transfusion; blood; haemolysis; infusion pump; pRBC.
Conflict of interest statement
The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Similar articles
-
In vitro quality control analysis after processing and during storage of feline packed red blood cells units.BMC Vet Res. 2018 Apr 27;14(1):141. doi: 10.1186/s12917-018-1458-4. BMC Vet Res. 2018. PMID: 29699565 Free PMC article.
-
In vitro hemolysis of stored units of canine packed red blood cells.J Vet Emerg Crit Care (San Antonio). 2018 Nov;28(6):512-517. doi: 10.1111/vec.12770. Epub 2018 Oct 9. J Vet Emerg Crit Care (San Antonio). 2018. PMID: 30299571
-
Quantitative assessment of haemolysis secondary to modern infusion pumps.Vox Sang. 2017 Apr;112(3):201-209. doi: 10.1111/vox.12486. Epub 2017 Feb 15. Vox Sang. 2017. PMID: 28198026
-
The potential adverse effects of haemolysis.Blood Transfus. 2017 May;15(3):218-221. doi: 10.2450/2017.0311-16. Blood Transfus. 2017. PMID: 28518048 Free PMC article. Review.
-
Selection of anticoagulant-preservatives for canine and feline blood storage.Vet Clin North Am Small Anim Pract. 1995 Nov;25(6):1263-76. doi: 10.1016/s0195-5616(95)50153-6. Vet Clin North Am Small Anim Pract. 1995. PMID: 8619265 Review.
Cited by
-
2021 ISFM Consensus Guidelines on the Collection and Administration of Blood and Blood Products in Cats.J Feline Med Surg. 2021 May;23(5):410-432. doi: 10.1177/1098612X211007071. J Feline Med Surg. 2021. PMID: 33896248 Free PMC article.
References
-
- Poder TG, Boileau JC, Lafrenière R, et al.. Quantitative assessment of haemolysis secondary to modern infusion pumps. Vox Sang 2017; 112: 201–209. - PubMed
-
- Windberger U, Bartholovitsch A, Plasenzotti R, et al.. Whole blood viscosity, plasma viscosity and erythrocyte aggregation in nine mammalian species: reference values and comparison of data. Exp Physiol 2003; 88: 431–440. - PubMed
-
- Hughes J, McNaughton J, Andrews J, et al.. Infusion pump-mediated mechanical hemolysis in pediatric patients. Ann Clin Lab Sci 2015; 45: 140–147. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous