The use of micro-computed tomography to determine the accuracy of electronic working length with two apex locators
- PMID: 33731504
- DOI: 10.2334/josnusd.20-0466
The use of micro-computed tomography to determine the accuracy of electronic working length with two apex locators
Abstract
Purpose: This study evaluated the precision of electronic working length by microcomputed tomography using two electronic apex locators (EALs).
Methods: Twenty single-rooted permanent teeth without caries or restorations were selected as the subject teeth. The positions of the minor apical constriction (AC) and major apical foramen (AF) were measured by electronic root canal length, and microcomputed tomography was performed with the file inserted and fixed in the root canal. All teeth were measured individually and independently by two operators. The Mann-Whitney U-test was used to statistically test the AC and AF values using two EALs; P < 0.05 was defined as statistically significant.
Results: This was 65.0% within 1.5 mm in the case of two EALs on AC. This was more than 90.0% within 1.0 mm in the case of two EALs on AF. Comparison of the differences between the respective AC and AF of the measurements obtained using the two EALs revealed no significant difference.
Conclusion: The two EALs are devices that can greatly improve the accuracy of WL control.
Keywords: cementodentinal junction; electronic apex locator; microcomputed tomography; working length.
Similar articles
-
The Use of Micro-Computed Tomography to Determine the Accuracy of 2 Electronic Apex Locators and Anatomic Variations Affecting Their Precision.J Endod. 2016 Aug;42(8):1263-7. doi: 10.1016/j.joen.2016.04.024. Epub 2016 Jun 9. J Endod. 2016. PMID: 27291503
-
A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars.J Endod. 2018 Dec;44(12):1872-1877. doi: 10.1016/j.joen.2018.09.001. Epub 2018 Nov 1. J Endod. 2018. PMID: 30390968
-
Micro-CT Study of the In Vivo Accuracy of a Wireless Electronic Apex Locator.J Endod. 2022 Sep;48(9):1152-1160. doi: 10.1016/j.joen.2022.06.003. Epub 2022 Jun 30. J Endod. 2022. PMID: 35780927
-
Determination of working length for teeth with wide or immature apices: a review.Int Endod J. 2013 Jun;46(6):483-91. doi: 10.1111/iej.12032. Epub 2012 Dec 7. Int Endod J. 2013. PMID: 23216284 Review.
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
Cited by
-
Evaluation of changes in root canal length and accuracy of the electronic apex locator during different stages of endodontic treatment and retreatment.Odontology. 2024 Apr;112(2):537-545. doi: 10.1007/s10266-023-00846-5. Epub 2023 Aug 29. Odontology. 2024. PMID: 37644294
-
Accuracy of apical limit control during retreatment provided by hybrid electronic foraminal locators: A micro-CT study.J Clin Exp Dent. 2025 Jan 1;17(1):e64-e69. doi: 10.4317/jced.62405. eCollection 2025 Jan. J Clin Exp Dent. 2025. PMID: 39958244 Free PMC article.
-
Evaluation of the effects of different file systems and apical functions of integrated endodontic motors on debris extrusion: an ex vivo experimental study.Restor Dent Endod. 2025 May;50(2):e14. doi: 10.5395/rde.2025.50.e14. Epub 2025 Apr 14. Restor Dent Endod. 2025. PMID: 40350563 Free PMC article.
-
The evaluation of the accuracy of a wireless electronic apex locator in primary molar teeth.BMC Oral Health. 2024 Dec 31;24(1):1580. doi: 10.1186/s12903-024-05387-z. BMC Oral Health. 2024. PMID: 39741258 Free PMC article.
-
Influence of root canal taper on the measurement of two different electronic apex locators.J Clin Exp Dent. 2024 Jun 1;16(6):e733-e739. doi: 10.4317/jced.61352. eCollection 2024 Jun. J Clin Exp Dent. 2024. PMID: 39130363 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous