Transorally obtained oxygen tension as an indicator of arterial oxygen tension
- PMID: 3373255
- DOI: 10.1007/BF01641807
Transorally obtained oxygen tension as an indicator of arterial oxygen tension
Abstract
Transcutaneous oxygen electrodes have been used with success in neonates as indicators of arterial oxygenation, but with less success in adults because of differences in skin thickness and vascularity. In this study, a prototype transoral oxygen electrode was evaluated to determine if a heated mucous membrane would yield arterialized values of oxygen tension in adults. Using a miniaturized Clark electrode, we measured transoral oxygen tension (PtoO2) in 29 subjects at steady-state conditions. Simultaneously a sample was anaerobically obtained from a radial artery for measurement of arterial oxygen tension (PaO2). Data were analyzed using linear regression analysis, Student's t test, and analysis of variance. There was no statistically significant difference between nonwhite and white subjects or male and female subjects. There was a highly significant difference (P less than 0.001) between the pooled, matched values for PtoO2 versus PaO2, and the regression between the PtoO2 and the PaO2 was linear (slope 0.92, y-intercept -8.37, r = 0.62, P less than 0.003). The calculated ratio of PtoO2 to PaO2 was 0.83 +/- 0.03 (standard error). We concluded that the PtoO2 was linearly related to the PaO2, although its accuracy in reflecting PaO2 was low. This finding correlates with previously published data that suggested that the PtoO2 reflects tissue oxygen tension rather than arterialized oxygen tension. Gender and race appeared not to affect the function of the electrode in our study.
Similar articles
-
Conjunctival oxygen tension and its relationship to arterial oxygen tension.J Clin Monit. 1986 Apr;2(2):100-4. doi: 10.1007/BF01637676. J Clin Monit. 1986. PMID: 3086511
-
Estimation of arterial oxygen tension in adult subjects using a transcutaneous electrode.Thorax. 1981 Jun;36(6):473-7. doi: 10.1136/thx.36.6.473. Thorax. 1981. PMID: 7314020 Free PMC article.
-
Continuous comparison of transcutaneous and arterial oxygen tension in newborn infants with respiratory illnesses.Acta Anaesthesiol Scand Suppl. 1978;68:91-7. doi: 10.1111/j.1399-6576.1978.tb01400.x. Acta Anaesthesiol Scand Suppl. 1978. PMID: 279208
-
Blood flow rate, temperature, oxygen tension and consumption in the skin of adults measured by a heated microcathode oxygen electrode.Dan Med Bull. 1988 Aug;35(4):322-34. Dan Med Bull. 1988. PMID: 3048920 Review.
-
Transcutaneous PO2 measurement.Can Anaesth Soc J. 1984 Nov;31(6):664-77. doi: 10.1007/BF03008765. Can Anaesth Soc J. 1984. PMID: 6388743 Review.