Measurement of right atrial oxygen saturation by fiberoptic oximetry accurately reflects mixed venous oxygen saturation in swine
- PMID: 3373256
- DOI: 10.1007/BF01641809
Measurement of right atrial oxygen saturation by fiberoptic oximetry accurately reflects mixed venous oxygen saturation in swine
Abstract
Continuous fiberoptic measurement of mixed venous oxygen saturation (SvO2) via a pulmonary artery catheter is a useful, though invasive, monitoring technique. Continuous right atrial venous oxygen saturation measurement by oximetry offers the potential of a significantly less invasive SvO2 measurement. However, catheter motion, character of the vessel, chamber wall reflection, the filtering technique involved in calculating oxygen saturation, and the streaming of venous blood prior to ventricular mixing may influence the feasibility of continuous right atrial (RA) SvO2 measurement. This study investigated the performance of fiberoptically measured RA SvO2, at a position 2 cm from the tricuspid valve, relative to simultaneously measured pulmonary artery (PA) SvO2. Ten pigs were subjected to circulatory shock or chemically induced lung damage. Over a total monitoring period of approximately 40 hours, 464 paired data points were sampled at 5-minute intervals. The difference between the overall means of RA and PA SvO2 was 0.91% with a standard error of the estimate of 4.7%, a regression equation of RA SvO2 = PA SvO2 (0.94 + 2.1) PA So2, and a correlation coefficient of 0.94. Our conclusion, although extrapolated from a pig model, is that fiberoptic SvO2 monitoring may be accomplished less invasively and at a lower cost with a right atrial catheter.
Similar articles
-
Capnodynamic assessment of mixed venous oxygen saturation in a porcine experimental endotoxemic model.Sci Rep. 2024 Nov 5;14(1):26807. doi: 10.1038/s41598-024-77483-7. Sci Rep. 2024. PMID: 39500994 Free PMC article.
-
Continuous mixed venous oxygen saturation measurement: a significant advance in hemodynamic monitoring?J Clin Monit. 1985 Oct;1(4):245-58. doi: 10.1007/BF02832818. J Clin Monit. 1985. PMID: 3938481 Review.
-
Clinical experience with continuous monitoring of mixed venous oxygen saturation in respiratory failure.Chest. 1984 Nov;86(5):748-52. doi: 10.1378/chest.86.5.748. Chest. 1984. PMID: 6488914
-
[Continuous monitoring of critical patients with a newly developed pulmonary arterial catheter. A cost analysis].Anaesthesist. 1995 Jun;44(6):423-8. doi: 10.1007/s001010050171. Anaesthesist. 1995. PMID: 7653794 Clinical Trial. German.
-
Continuous cardiac output and mixed venous oxygen saturation monitoring.J Crit Care. 1998 Dec;13(4):198-213. doi: 10.1016/s0883-9441(98)90006-5. J Crit Care. 1998. PMID: 9869547 Review.
Cited by
-
Continuous central venous oxygen saturation (ScvO2) measurement using a fibre optic catheter in newborn infants.Arch Dis Child Fetal Neonatal Ed. 1996 May;74(3):F177-81. doi: 10.1136/fn.74.3.f177. Arch Dis Child Fetal Neonatal Ed. 1996. PMID: 8777680 Free PMC article.
-
Incidence of low central venous oxygen saturation during unplanned admissions in a multidisciplinary intensive care unit: an observational study.Crit Care. 2007;11(1):R2. doi: 10.1186/cc5144. Crit Care. 2007. PMID: 17212816 Free PMC article.
-
Hemodynamic Monitoring in the Acute Management of Pediatric Heart Failure.Curr Cardiol Rev. 2016;12(2):112-6. doi: 10.2174/1573403x12666151119165007. Curr Cardiol Rev. 2016. PMID: 26585037 Free PMC article.
-
Superior vena caval and mixed venous oxyhemoglobin saturations in children recovering from open heart surgery.J Clin Monit. 1992 Jan;8(1):44-9. doi: 10.1007/BF01618087. J Clin Monit. 1992. PMID: 1538252