A systematic evaluation of medical 3D printing accuracy of multi-pathological anatomical models for surgical planning manufactured in elastic and rigid material using desktop inverted vat photopolymerization
- PMID: 33733499
- DOI: 10.1002/mp.14850
A systematic evaluation of medical 3D printing accuracy of multi-pathological anatomical models for surgical planning manufactured in elastic and rigid material using desktop inverted vat photopolymerization
Abstract
Purpose: The dimensional accuracy of three-dimensional (3D) printed anatomical models is essential to correctly understand spatial relationships and enable safe presurgical planning. Most recent accuracy studies focused on 3D printing of a single pathology for surgical planning. This study evaluated the accuracy of medical models across multiple pathologies, using desktop inverted vat photopolymerization (VP) to 3D print anatomic models using both rigid and elastic materials.
Methods: In the primary study, we 3D printed seven models (six anatomic models and one reference cube) with volumes ranging from ~2 to ~209 cc. The anatomic models spanned multiple pathologies (neurological, cardiovascular, abdominal, musculoskeletal). Two solid measurement landing blocks were strategically created around the pathology to allow high-resolution measurement using a digital micrometer and/or caliper. The physical measurements were compared to the designed dimensions, and further analysis was conducted regarding the observed patterns in accuracy. All of the models were printed in three resins: Elastic, Clear, and Grey Pro in the primary experiments. A full factorial block experimental design was employed and a total of 42 models were 3D printed in 21 print runs. In the secondary study, we 3D printed two of the anatomic models in triplicates selected from the previous six to evaluate the effect of 0.1 mm vs 0.05 mm layer height on the accuracy.
Results: In the primary experiment, all dimensional errors were less than 1 mm. The average dimensional error across the 42 models was 0.238 0.219 mm and the relative error was 1.10 1.13%. Results from the secondary experiments were similar with an average dimensional error of 0.252 0.213 mm and relative error of 1.52% 1.28% across 18 models. There was a statistically significant difference in the relative errors between the Elastic resin and Clear resin groups. We explained this difference by evaluating inverted VP 3D printing peel forces. There was a significant difference between the Solid and Hollow group of models. There was a significant difference between measurement landing blocks oriented Horizontally and Vertically. In the secondary experiments, there was no difference in accuracy between the 0.10 and 0.05 mm layer heights.
Conclusions: The maximum measured error was less than 1 mm across all models, and the mean error was less than 0.26mm. Therefore, inverted VP 3D printing technology is suitable for medical 3D printing if 1 mm is considered the cutoff for clinical use cases. The 0.1 mm layer height is suitable for 3D printing accurate anatomical models for presurgical planning in a majority of cases. Elastic models, models oriented horizontally, and models that are hollow tend to have relatively higher deviation as seen from experimental results and mathematical model predictions. While clinically insignificant using a 1 mm cutoff, further research is needed to better understand the complex physical interactions in VP 3D printing which influence model accuracy.
Keywords: Form 3B; additive manufacturing; anatomic models; elastic material; inverted vat polymerization; medical 3D printing; model accuracy; rapid prototyping; vat photopolymerization.
© 2021 American Association of Physicists in Medicine.
Similar articles
-
Medical 3D Printing Dimensional Accuracy for Multi-pathological Anatomical Models 3D Printed Using Material Extrusion.J Digit Imaging. 2022 Jun;35(3):613-622. doi: 10.1007/s10278-022-00614-x. Epub 2022 Mar 2. J Digit Imaging. 2022. PMID: 35237891 Free PMC article.
-
Medical 3D Printing Using Desktop Inverted Vat Photopolymerization: Background, Clinical Applications, and Challenges.Bioengineering (Basel). 2023 Jun 30;10(7):782. doi: 10.3390/bioengineering10070782. Bioengineering (Basel). 2023. PMID: 37508810 Free PMC article. Review.
-
A focused simulation-based optimization of print time and material usage with respect to orientation, layer height and support settings for multi-pathological anatomical models in inverted vat photopolymerization 3D printing.3D Print Med. 2021 Aug 26;7(1):23. doi: 10.1186/s41205-021-00112-w. 3D Print Med. 2021. PMID: 34448082 Free PMC article.
-
Accuracy and feasibility in building a personalized 3D printed femoral pseudoaneurysm model for endovascular training.PLoS One. 2024 Jun 3;19(6):e0304506. doi: 10.1371/journal.pone.0304506. eCollection 2024. PLoS One. 2024. PMID: 38829913 Free PMC article.
-
Accuracy of 3D Printed Models Created by Two Technologies of Printers with Different Designs of Model Base.J Prosthodont. 2020 Feb;29(2):124-128. doi: 10.1111/jopr.13107. Epub 2019 Sep 9. J Prosthodont. 2020. PMID: 31498957 Review.
Cited by
-
Medical 3D Printing Dimensional Accuracy for Multi-pathological Anatomical Models 3D Printed Using Material Extrusion.J Digit Imaging. 2022 Jun;35(3):613-622. doi: 10.1007/s10278-022-00614-x. Epub 2022 Mar 2. J Digit Imaging. 2022. PMID: 35237891 Free PMC article.
-
Graphene in 3D Bioprinting.J Funct Biomater. 2024 Mar 25;15(4):82. doi: 10.3390/jfb15040082. J Funct Biomater. 2024. PMID: 38667539 Free PMC article. Review.
-
The biomechanical behavior of 3D printed human femoral bones based on generic and patient-specific geometries.3D Print Med. 2022 Nov 23;8(1):35. doi: 10.1186/s41205-022-00162-8. 3D Print Med. 2022. PMID: 36418789 Free PMC article.
-
The Role of 3D Printing in Planning Complex Medical Procedures and Training of Medical Professionals-Cross-Sectional Multispecialty Review.Int J Environ Res Public Health. 2022 Mar 11;19(6):3331. doi: 10.3390/ijerph19063331. Int J Environ Res Public Health. 2022. PMID: 35329016 Free PMC article. Review.
-
Medical 3D Printing Using Desktop Inverted Vat Photopolymerization: Background, Clinical Applications, and Challenges.Bioengineering (Basel). 2023 Jun 30;10(7):782. doi: 10.3390/bioengineering10070782. Bioengineering (Basel). 2023. PMID: 37508810 Free PMC article. Review.
References
REFERENCES
-
- Hazelaar C, Van Eijnatten M, Dahele M, et al. Using 3D printing techniques to create an anthropomorphic thorax phantom for medical imaging purposes. Med Phys. 2018;45:92-100.
-
- Tappa K, Jammalamadaka U. Novel biomaterials used in medical 3D printing techniques. J Funct Biomater. 2018;9:17.
-
- Ravi P, Wright J, Shiakolas PS, Welch TR. Three-dimensional printing of poly(glycerol sebacate fumarate) gadodiamide-poly(ethylene glycol) diacrylate structures and characterization of mechanical properties for soft tissue applications. J Biomed Mater Res Part B Appl Biomater. 2019;107:664-671.
-
- Zou W, Fisher T, Zhang M, et al. Potential of 3D printing technologies for fabrication of electron bolus and proton compensators. J Appl Clin Med Phys. 2015;16:90-98.
-
- Filippou V, Tsoumpas C. Recent advances on the development of phantoms using 3D printing for imaging with CT, MRI, PET, SPECT, and ultrasound. Med Phys. 2018;45:e740-e760.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources