Randomized Controlled Trial
. 2021 May;41(3):273-283.
doi: 10.1177/0896860821993954. Epub 2021 Mar 18.

Clinical effectiveness of a Malaysian-manufactured CAPD product: A randomised trial

Affiliations
Randomized Controlled Trial

Clinical effectiveness of a Malaysian-manufactured CAPD product: A randomised trial

Wen Yao Mak et al. Perit Dial Int. 2021 May.

Abstract

Background: We compared the clinical effectiveness of a new peritoneal dialysis (PD) product with polyvinyl chloride-containing tubing (Stay Safe Link®, SSL) with the plastic-free alternative (Stay Safe®, STS) in patients on continuous ambulatory peritoneal dialysis (CAPD).

Method: A multicentre, parallel, randomised, controlled, open-label, non-inferiority trial was conducted. Adult patients receiving CAPD were randomised in a 1:1 ratio to SSL or STS. The primary outcome was the rate of peritonitis after 1 year of follow-up.

Results: A total of 472 subjects were randomised (SSL, n = 233; STS, n = 239). One subject in each group was excluded from the analysis as they withdrew consent before the first dialysis dose. Four hundred and seventy subjects (SSL, n = 232; STS, n = 238) were included in the modified intention-to-treat analysis. Non-inferiority between two groups was established as no significant difference was found in peritonitis rate (incident rate ratio: 0.91, 95% CI: 0.65-1.28). No significant difference was detected in weekly Kt/V (p = 0.58) and creatinine clearance (p = 0.55). However, the average ultrafiltration volume was significantly lower in SSL, with a mean difference of 93 ml (p < 0.01). SSL also demonstrated a 2.57-times higher risk of device defect than STS (95% CI: 1.77-3.75).

Conclusion: SSL was non-inferior in peritonitis rate compared to plastic-free STS over 1 year in patients requiring CAPD. There was no difference in the delivered dialysis dose, but there was a higher rate of device defects with SSL.

Keywords: Creatinine clearance; Kt/V; device defect; peritoneal dialysis; peritonitis; plasticiser; ultrafiltration rate.