Comparative Study
. 1979 Jan 20;1(6157):156-8.
doi: 10.1136/bmj.1.6157.156.

Comparative study of automatic blood-gas analysers and their use in analysing arterial and capillary samples

Comparative Study

Comparative study of automatic blood-gas analysers and their use in analysing arterial and capillary samples

P Rubin et al. Br Med J. .

Abstract

Three automatic blood-gas analysers were compared for ease of use; calibration; reproducibility and accuracy of results; maintenance; fault-finding; and use of expert technician time. Results obtained from arterial and capillary blood were compared with duplicate values obtained with a semi-automatic analyser controlled and calibrated with tonometered blood. No analyser was fully automatic, and all three needed maintenance by expert technicians. Difficulties were encountered when inexperienced operators used the machines. One automatic blood-gas analyser gave aberrant values for oxygen pressure (PO2) due to electrode dysfunction that was not indicated by the fault-finding system. A second analyser gave significantly lower values for blood pH than the standard machine. A comparison of pH, carbon dioxide pressure (PCO2), and PO2 measured in 40 simultaneous paired samples of arterial and arterialised capillary blood showed no significant difference for pH or PCO2, but the PO2 values were significantly lower in the capillary samples over the range studied. We conclude that all machines perform satisfactorily in terms of blood-gas analysis, but none may be regarded as fully automatic. Some degree of technical supervision is essential, as is proper training for all potential users.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Br J Dis Chest. 1976 Oct;70(4):263-8 - PubMed
    1. Ann Clin Biochem. 1977 Sep;14(5):245-53 - PubMed
    1. Br J Anaesth. 1974 Apr;46(4):253-9 - PubMed
    1. Nature. 1972 Oct 20;239(5373):474 - PubMed
    1. Br Med J. 1968 Sep 14;3(5619):640-3 - PubMed

Publication types

LinkOut - more resources