[Treatment of ventricular tachyarrhythmias with an implantable cardioverter-defibrillator system]
- PMID: 3376474
[Treatment of ventricular tachyarrhythmias with an implantable cardioverter-defibrillator system]
Abstract
5 patients (3 with coronary artery disease and chronic myocardial infarction, 2 with dilatative cardiomyopathy) with a mean age of 59 years (range 54-69 years) with drug refractory ventricular tachycardia and/or ventricular fibrillation received the automatic implantable cardioverter defibrillator (AICD). Intraoperative testing revealed a mean defibrillation threshold of 13 +/- 2.7 Joule. Over a mean follow-up period of 15.2 months (range 3-25 months) the patients received a total of 117 discharges. 15% of the delivered shocks were recorded during continuous ECG monitoring, 13% were associated with palpitations and 27% were discharged during syncope. 45% of shocks occurred in the absence of symptoms. No patient died suddenly. 1 patient died of intractable heart failure, 1 patient died of septic shock. In carefully selected patients the automatic implantable cardioverter defibrillator is an effective tool in the treatment of life-threatening ventricular tachyarrhythmias. Modifications of the device to incorporate programmability of the cut-off rate, the sensing criteria and the levels of shock energy, as well as the options for different pacing modes combined with memory functions are needed to improve antiarrhythmic strategies.