. 2021 Mar;8(1):e000739.
doi: 10.1136/bmjresp-2020-000739.

In vitro evaluation of disposable transport ventilators with combination aerosol therapy

Affiliations

In vitro evaluation of disposable transport ventilators with combination aerosol therapy

Marc Mac Giolla Eain et al. BMJ Open Respir Res. 2021 Mar.

Abstract

Background: The COVID-19 pandemic has highlighted the need for alternative short-term, reliable means to aid in the treatment of patients requiring ventilatory support. Concurrent aerosol drug delivery is often prescribed to such patients. As such, this study examines one such short-term option, the disposable gas-powered transport ventilator to effectively deliver aerosol therapy. Factors such as aerosol generator type, patient breathing pattern, humidification and nebuliser position within the respiratory circuit were also examined.

Methods: Aerosol drug delivery characterisation was undertaken using two different disposable transport ventilators (DTVs). Two different nebuliser types, a closed circuit vibrating mesh nebuliser (VMN) and an open circuit jet nebuliser (JN), at different locations in a respiratory circuit, proximal and distal to an endotracheal tube (ETT), with and without passive humidification, were evaluated in simulated adult and paediatric patients.

Results: Placement of a nebuliser proximal to the ETT (VMN: 25.19%-34.15% and JN: 3.14%-8.92%), and the addition of a heat and moisture exchange filter (VMN: 32.37%-40.43% and JN: 5.60%-9.91%) resulted in the largest potential lung dose in the adult patient model. Irrespective of nebuliser position and humidification in the respiratory circuit, use of the VMN resulted in the largest potential lung dose (%). A similar trend was recorded in the paediatric model data, where the largest potential lung dose was recorded with both nebuliser types placed proximal to the ETT (VMN: 8.12%-10.89% and JN: 2.15%-3.82%). However, the addition of a heat and moisture exchange filter had no statistically significant effect on the potential lung dose (%) a paediatric patient would receive (p>>0.05).

Conclusions: This study demonstrates that transport ventilators, such as DTVs, can be used concurrently with aerosol generators to effectively deliver aerosolised medication in both adult and paediatric patients.

Keywords: ambulatory oxygen therapy; assisted ventilation; nebuliser therapy.

PubMed Disclaimer

Conflict of interest statement

Competing interests: All authors are employees of Aerogen Limited.

Figures

Figure 1
Figure 1
Schematic illustration of the test set-up. Simulated adult and paediatric ventilation assessed the lung dose (%) beyond the ETT across two nebuliser positions: (i) proximal and (ii) distal to the ETT. ETT, endotracheal tube.
Figure 2
Figure 2
Box plot presentation of the variations in lung dose (%) with changes in nebuliser, passive humidification and nebuliser position within the respiratory circuit for a simulated intubated adult patient. The plots display the minimum, first quartile, median, third quartile and maximum lung dosages (%) measured. HME, heat and moisture exchanger; JN, jet nebuliser; VMN, vibrating mesh nebuliser.
Figure 3
Figure 3
Box plot presentation of the variations in lung dose (%) with changes in nebuliser, passive humidification and nebuliser position within the respiratory circuit for a simulated intubated paediatric patient. The plots display the minimum, first quartile, median, third quartile and maximum lung dosages (%) measured. HME, heat and moisture exchanger; JN, jet nebuliser; VMN, vibrating mesh nebuliser.

Similar articles

Cited by

References

    1. Nates JL. Combined external and internal Hospital disaster: impact and response in a Houston trauma center intensive care unit. Crit Care Med 2004;32:686–90. 10.1097/01.CCM.0000114995.14120.6D - DOI - PubMed
    1. L'Her E, Roy A, Marjanovic N. Bench-test comparison of 26 emergency and transport ventilators. Crit Care 2014;18:1–14. 10.1186/s13054-014-0506-0 - DOI - PMC - PubMed
    1. L'Her E, Roy A, L’Her E. Bench tests of simple, handy ventilators for pandemics: performance, autonomy, and ergonomy. Respir Care 2011;56:751–60. 10.4187/respcare.00841 - DOI - PubMed
    1. Terms F. Evaluation of resuscitators. J Med Eng Technol 1984;8:71–8. 10.3109/03091908409036747 - DOI - PubMed
    1. Weiss SJ, Ernst AA, Jones R, et al. . Automatic transport ventilator versus bag valve in the EMS setting: a prospective, randomized trial. South Med J 2005;98:970–6. 10.1097/01.smj.0000182177.01436.70 - DOI - PubMed