In vitro evaluation of disposable transport ventilators with combination aerosol therapy
- PMID: 33771812
- PMCID: PMC8006849
- DOI: 10.1136/bmjresp-2020-000739
In vitro evaluation of disposable transport ventilators with combination aerosol therapy
Abstract
Background: The COVID-19 pandemic has highlighted the need for alternative short-term, reliable means to aid in the treatment of patients requiring ventilatory support. Concurrent aerosol drug delivery is often prescribed to such patients. As such, this study examines one such short-term option, the disposable gas-powered transport ventilator to effectively deliver aerosol therapy. Factors such as aerosol generator type, patient breathing pattern, humidification and nebuliser position within the respiratory circuit were also examined.
Methods: Aerosol drug delivery characterisation was undertaken using two different disposable transport ventilators (DTVs). Two different nebuliser types, a closed circuit vibrating mesh nebuliser (VMN) and an open circuit jet nebuliser (JN), at different locations in a respiratory circuit, proximal and distal to an endotracheal tube (ETT), with and without passive humidification, were evaluated in simulated adult and paediatric patients.
Results: Placement of a nebuliser proximal to the ETT (VMN: 25.19%-34.15% and JN: 3.14%-8.92%), and the addition of a heat and moisture exchange filter (VMN: 32.37%-40.43% and JN: 5.60%-9.91%) resulted in the largest potential lung dose in the adult patient model. Irrespective of nebuliser position and humidification in the respiratory circuit, use of the VMN resulted in the largest potential lung dose (%). A similar trend was recorded in the paediatric model data, where the largest potential lung dose was recorded with both nebuliser types placed proximal to the ETT (VMN: 8.12%-10.89% and JN: 2.15%-3.82%). However, the addition of a heat and moisture exchange filter had no statistically significant effect on the potential lung dose (%) a paediatric patient would receive (p>>0.05).
Conclusions: This study demonstrates that transport ventilators, such as DTVs, can be used concurrently with aerosol generators to effectively deliver aerosolised medication in both adult and paediatric patients.
Keywords: ambulatory oxygen therapy; assisted ventilation; nebuliser therapy.
© Author(s) (or their employer(s)) 2021. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: All authors are employees of Aerogen Limited.
Figures



Similar articles
-
An in vitro study of the effects of respiratory circuit setup and parameters on aerosol delivery during mechanical ventilation.Front Med (Lausanne). 2024 Jan 24;10:1307301. doi: 10.3389/fmed.2023.1307301. eCollection 2023. Front Med (Lausanne). 2024. PMID: 38327274 Free PMC article.
-
An in vitro visual study of fugitive aerosols released during aerosol therapy to an invasively ventilated simulated patient.Drug Deliv. 2021 Dec;28(1):1496-1500. doi: 10.1080/10717544.2021.1951893. Drug Deliv. 2021. PMID: 34259091 Free PMC article.
-
Clinical outcome associated with the use of different inhalation method with and without humidification in asthmatic mechanically ventilated patients.Pulm Pharmacol Ther. 2017 Aug;45:40-46. doi: 10.1016/j.pupt.2017.04.007. Epub 2017 Apr 20. Pulm Pharmacol Ther. 2017. PMID: 28435031 Clinical Trial.
-
Not all nebulizers are created equal: Considerations in choosing a nebulizer for aerosol delivery during mechanical ventilation.Expert Rev Respir Med. 2023 Feb;17(2):131-142. doi: 10.1080/17476348.2023.2183194. Epub 2023 Feb 27. Expert Rev Respir Med. 2023. PMID: 36803134 Review.
-
How best to deliver aerosol medications to mechanically ventilated patients.Clin Chest Med. 2008 Jun;29(2):277-96, vi. doi: 10.1016/j.ccm.2008.02.003. Clin Chest Med. 2008. PMID: 18440437 Review.
Cited by
-
The Impact of Head Model Choice on the In Vitro Evaluation of Aerosol Drug Delivery.Pharmaceutics. 2021 Dec 23;14(1):24. doi: 10.3390/pharmaceutics14010024. Pharmaceutics. 2021. PMID: 35056920 Free PMC article.
-
Cellular Therapy for the Treatment of Paediatric Respiratory Disease.Int J Mol Sci. 2021 Aug 18;22(16):8906. doi: 10.3390/ijms22168906. Int J Mol Sci. 2021. PMID: 34445609 Free PMC article. Review.
-
An in vitro study of the effects of respiratory circuit setup and parameters on aerosol delivery during mechanical ventilation.Front Med (Lausanne). 2024 Jan 24;10:1307301. doi: 10.3389/fmed.2023.1307301. eCollection 2023. Front Med (Lausanne). 2024. PMID: 38327274 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials