. 2021 Sep 1;56(9):545-552.
doi: 10.1097/RLI.0000000000000778.

Controlling Through-Slice Chemical-Shift Artifacts for Improved Non-Fat-Suppressed Musculoskeletal Turbo-Spin-Echo Magnetic Resonance Imaging at 7 T

Affiliations

Controlling Through-Slice Chemical-Shift Artifacts for Improved Non-Fat-Suppressed Musculoskeletal Turbo-Spin-Echo Magnetic Resonance Imaging at 7 T

Constantin von Deuster et al. Invest Radiol. .

Abstract

Objectives: Through-slice chemical shift artifacts in state-of-the-art turbo-spin-echo (TSE) images can be significantly more severe at 7 T than at lower field strengths. In musculoskeletal applications, these artifacts appear similar to bone fractures or neoplastic bone marrow disease. The objective of this work was to explore and reduce through-slice chemical shift artifacts in 2-dimensional (2D) TSE imaging at 7 T.

Materials and methods: This prospective study was approved by the local ethics board. The bandwidths of the excitation and refocusing radiofrequency (RF) pulses of a prototype 2D TSE sequence were individually modified and their effect on the slice profiles and relative slice locations of water and fat spins was assessed in an oil-water phantom. Based on these results, it was hypothesized that the combination of matched and increased excitation and refocusing RF pulse bandwidths ("MIB") of 1500 Hz would enable 2D TSE imaging with significantly reduced chemical shift artifacts compared with a state-of-the-art sequence with unmatched and moderate RF pulse bandwidths ("UMB") of 1095 and 682 Hz.A series of T1-weighted sagittal knee examinations in 10 healthy human subjects were acquired using the MIB and UMB sequences and independently evaluated by 2 radiologists. They measured the width of chemical shift artifacts at 2 standardized locations and graded the perceived negative effect of chemical shift artifacts on image quality in the bones and in the whole gastrocnemius muscle on a 5-point scale. Similar knee, wrist, and foot images were acquired in a single subject. Signal-to-noise ratios in the femoral bone marrow were computed between the UMB and MIB sequences.

Results: Phantom measurements confirmed the expected spatial separation of simultaneously affected water and fat slices between 40% and 200% of the prescribed slice thickness for RF pulse bandwidths between 2500 and 500 Hz. Through-slice chemical shift artifacts at the bone-cartilage interface were significantly smaller with MIB than with UMB (location 1: 0.35 ± 0.20 mm vs 1.27 ± 0.27 mm, P < 0.001; location 2: 0.25 ± 0.13 mm vs 1.48 ± 0.46 mm, P < 0.001; intraclass correlation coefficient = 0.98). The negative effect of chemical shift artifacts on image quality was significantly smaller with MIB than with UMB (bone: 2 ± 0 vs 4 ± 1, P < 0.004 [both readers]; muscle: 3 ± 0 vs 2 ± 0, P < 0.004 [both readers]; κ = 0.69). The signal-to-noise ratio of the UMB and MIB sequences was comparable, with a ratio of 99 ± 7%. Images acquired using the UMB sequence displayed numerous artifactual hyperintensities and diffuse, as well as locally severe, fat signal loss in all examined regions, whereas the MIB sequence consistently yielded high image quality with bright T1-weighted fat signal and excellent depiction of fine tissue structures.

Conclusions: On 7 T systems, the selection of high and matched RF bandwidths for excitation and refocusing pulses for 2D TSE imaging without fat suppression showed consistently better image quality than state-of-the-art sequences with unmatched lower RF pulse bandwidths.

PubMed Disclaimer

Conflict of interest statement

Conflicts of interest and sources of funding: Balgrist University Hospital of Zurich as well as Balgrist Campus AG each has a research agreement with Siemens Healthcare AG. C.v.D., S.S., and R.M.H. are employees of Siemens Healthcare AG.

References

    1. Krug R, Stehling C, Kelley DA, et al. Imaging of the musculoskeletal system in vivo using ultra-high field magnetic resonance at 7 T. Invest Radiol . 2009;44:613–618.
    1. Stahl R, Krug R, Kelley DAC, et al. Assessment of cartilage-dedicated sequences at ultra-high-field MRI: comparison of imaging performance and diagnostic confidence between 3.0 and 7.0 T with respect to osteoarthritis-induced changes at the knee joint. Skeletal Radiol . 2009;38:771–783.
    1. Welsch GH, Juras V, Szomolanyi P, et al. Magnetic resonance imaging of the knee at 3 and 7 tesla: a comparison using dedicated multi-channel coils and optimised 2D and 3D protocols. Eur Radiol . 2012;22:1852–1859.
    1. Lazik A, Theysohn JM, Geis C, et al. 7 Tesla quantitative hip MRI: T1, T2 and T2* mapping of hip cartilage in healthy volunteers. Eur Radiol . 2016;26:1245–1253.
    1. Wyatt C, Guha A, Venkatachari A, et al. Improved differentiation between knees with cartilage lesions and controls using 7T relaxation time mapping. J Orthop Translat . 2015;3:197–204.

Publication types

LinkOut - more resources