[Development of a System for Evaluating the In-room-laser Alignment Including the Horizontality and Verticality Integrated the Light/Radiation Field Coincidence Test and the Winston-Lutz Test]
- PMID: 33853977
- DOI: 10.11323/jjmp.41.1_1
[Development of a System for Evaluating the In-room-laser Alignment Including the Horizontality and Verticality Integrated the Light/Radiation Field Coincidence Test and the Winston-Lutz Test]
Abstract
Purpose: The in-room laser which is used for patient positioning in radiotherapy is generally projected on the radiation isocenter determined by the Winston-Lutz test and so on. In this study, a couch-mounted verification device was developed that could evaluate all in-room lasers' alignment including the horizontality and verticality at one time. The device has the function to perform the light/radiation field coincidence test and the Winston-Lutz test at the same time. The aim of this report was to introduce the verification procedure for two tests, using the newly developed software and device, and to present the tuning flow of the in-room laser. Moreover, the analysis accuracy of the developed software was evaluated in comparison with commercial software.
Methods: First, the light/radiation field was evaluated by using tungsten markers on the central surface of the device. Next, after aligning the long-carved lines on the front and sides of the device with the in-room lasers, the Winston-Lutz test was carried out by using the tungsten sphere in the center of the device. The acquired images were collectively analyzed using the developed software equipped with the reporting function. Additionally, the result of this Winston-Lutz test was compared with the result from commercial software.
Results: A series of the light/radiation field coincidence test and the Winston-Lutz test were analyzed using the developed device and software. The results could be easily confirmed using the reporting function of the software. Regarding the result of the Winston-Lutz test, most of the analysis differences between the developed software and commercially available software were within the pixel size (0.22 mm).
Discussions: Since the accuracy of the radiation field affects the result of the Winston-Lutz test, the presented procedure of performing the light/radiation coincidence field test in advance facilitates the interpretation of the error of the Winston-Lutz test. Based on the results of the Winston-Lutz test, we were able to demonstrate the tuning flow of all in-room lasers including the horizontality and verticality by using the developed device.
Conclusions: We have developed a couch-mounted verification device and software that can evaluate the light/radiation coincidence field test and the alignment including the horizontality and verticality of the in-room laser used for patient positioning in radiotherapy, and reported its usefulness. The analysis accuracy of the developed software was comparable to that of commercially available software. The use of this device and the developed software would contribute to not only the efficiency of adjusting all in-room lasers' alignment including the horizontality and verticality but also reflect accurately the result of the Winston-Lutz test.
Keywords: laser, Winston-Lutz test; light field; quality control; radiation field.
Similar articles
-
Off-iso Winston-Lutz test on seven linear accelerators.J Appl Clin Med Phys. 2024 Oct;25(10):e14470. doi: 10.1002/acm2.14470. Epub 2024 Jul 23. J Appl Clin Med Phys. 2024. PMID: 39042435 Free PMC article.
-
Feasibility of virtual starshot analysis providing submillimeter radiation isocenter accuracy: A long-term multi-institutional analysis.J Appl Clin Med Phys. 2019 Oct;20(10):74-83. doi: 10.1002/acm2.12715. Epub 2019 Sep 9. J Appl Clin Med Phys. 2019. PMID: 31502408 Free PMC article.
-
Positions of radiation isocenter and the couch rotation center established by Winston-Lutz and optical measurements.Tech Innov Patient Support Radiat Oncol. 2022 Feb 17;21:46-50. doi: 10.1016/j.tipsro.2022.01.004. eCollection 2022 Mar. Tech Innov Patient Support Radiat Oncol. 2022. PMID: 35243044 Free PMC article.
-
A study of Winston-Lutz test on two different electronic portal imaging devices and with low energy imaging.Australas Phys Eng Sci Med. 2016 Sep;39(3):677-85. doi: 10.1007/s13246-016-0463-9. Epub 2016 Jul 19. Australas Phys Eng Sci Med. 2016. PMID: 27435984
-
Portal dosimetry scripting application programming interface (PDSAPI) for Winston-Lutz test employing ceramic balls.J Appl Clin Med Phys. 2020 Nov;21(11):295-303. doi: 10.1002/acm2.13043. Epub 2020 Oct 24. J Appl Clin Med Phys. 2020. PMID: 33098369 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources