[Optimal Beam Quality in Chest Radiography Using CsI-flat Panel Detector for Detection of Pulmonary Nodules]
- PMID: 33883367
- DOI: 10.6009/jjrt.2021_JSRT_77.4.335
[Optimal Beam Quality in Chest Radiography Using CsI-flat Panel Detector for Detection of Pulmonary Nodules]
Abstract
Objectives: Optimal beam quality for detection of pulmonary nodules in digital chest radiography using CsI-flat panel detector (FPD) was investigated in consideration of image quality and patient dose.
Methods: The human chest phantom with inserted imitated nodules (diameter: 10 mm, CT value: +30 Hounsfield unit (HU), -375 HU, -620 HU) was used for the measurement of contrast-to-noise ratio (CNR) of imitated nodules by twenty beams arranged by five tube voltages and four filters.
Results: The CNR varies with X-ray tube voltage and added filter. CNR correlates weakly to the tube voltage, fairly to the effective energy in second-order polynomial and strongly to the quality index (effective energy divided X-ray tube voltage). In order to improve the CNR, the effective energy and the quality index are kept about 50 keV and more than 0.5, respectively, using an 80-100 kV beam with a copper filter.
Conclusion: A 90 kV (2.5 mm Al inherent filtration) beam with a 0.15 mm copper filter and a 90 kV or 100 kV (2.5 mm Al inherent filtration) beam with a 0.2 mm copper filter are appropriate for chest radiography using CsI-FPD.
Keywords: CsI-flat panel detector (FPD); beam quality; contrast-to-noise ratio (CNR); digital chest radiography (DR); pulmonary nodule.
Similar articles
-
[Study of X-ray Beam Quality and Contrast-to-Noise Ratio of Lung Nodules in Chest FPD Radiography: Monte Carlo Simulation Using Chest Model Phantom].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2022;78(2):159-170. doi: 10.6009/jjrt.780206. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2022. PMID: 35185095 Japanese.
-
Optimal beam quality for chest flat panel detector system: realistic phantom study.Eur Radiol. 2019 Sep;29(9):4538-4543. doi: 10.1007/s00330-019-5998-1. Epub 2019 Feb 8. Eur Radiol. 2019. PMID: 30737566
-
[Evaluation of 90 kV Beam with 0.15 mm Cu Filter in Chest Radiography Using CsI-flat Panel Detector].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2020;76(5):463-473. doi: 10.6009/jjrt.2020_JSRT_76.5.463. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2020. PMID: 32435030 Japanese.
-
[Subjective Contrast of Pulmonary Tissues in CsI-FPD Chest Radiography Using Model Phantom by Monte Carlo Simulation].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2023 Feb 20;79(2):128-141. doi: 10.6009/jjrt.2023-1290. Epub 2023 Jan 18. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2023. PMID: 36653125 Japanese.
-
[Paradigm Shift in Respiratory Diagnosis: Current Status and Future Prospects of Dynamic Chest Radiography].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2021;77(11):1279-1287. doi: 10.6009/jjrt.2021_JSRT_77.11.1279. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2021. PMID: 34803108 Review. Japanese.
Cited by
-
Usefulness of copper filters in digital chest radiography based on the relationship between effective detective quantum efficiency and deep learning-based segmentation accuracy of the tumor area.Radiol Phys Technol. 2023 Jun;16(2):299-309. doi: 10.1007/s12194-023-00719-0. Epub 2023 Apr 12. Radiol Phys Technol. 2023. PMID: 37046154
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources