Abstract
Background: Cadmium-zinc-telluride (CZT) SPECT/CT cameras with large field of view offer a higher sensitivity than conventional Anger cameras. This prospective study aimed to determine the equivalence between a conventional protocol and a reduced acquisition time protocol for 201-Thallium myocardial perfusion imaging (MPI) using a whole-body CZT SPECT camera.
Methods and results: Stress MPI was obtained for 103 consecutive patients on a DISCOVERY-CZT camera. Images were anonymized and post-processed to simulate a 25% (D75 dataset) and 50% (D50 dataset) decrease in total recorded counts. Concerning the number of segments displaying a tracer uptake < 70% of maximum intensity per patient, equivalence was demonstrated for both count-reduced datasets with a good inter-observer agreement (between 0.90 and 0.88). When comparing the full-vs-D75 datasets and full-vs-D50 datasets, mean difference was 0.06 segment (CI95: [- 0.15;0.27], P < 0.001) and 0.518 segment (CI95: [0.28;0.76], P < 0.001) respectively. Inter-observer agreement was also moderate to good concerning the number of pathological segments (between 0.6 and 0.7) and excellent for functional parameters.
Conclusion: Whole-body CZT SPECT/CT cameras allow to reduce 201-Thallium MPI injected activity or acquisition time by 50% with an equivalence in the number of segments displaying a tracer uptake < 70% of maximum intensity and with a good inter-observer agreement.
Keywords: 201Thallium; Cadmium-zinc-telluride; SPECT; myocardial perfusion imaging.
© 2021. American Society of Nuclear Cardiology.
Comment in
-
Scintillation cameras: A new clinical era has come.J Nucl Cardiol. 2022 Aug;29(4):1942-1945. doi: 10.1007/s12350-021-02660-4. Epub 2021 Jul 12. J Nucl Cardiol. 2022. PMID: 34254249 No abstract available.
References
-
- Naghavi M, Wang H, Lozano R, et al. Global, regional, and national age-sex specific all-cause and cause-specific mortality for 240 causes of death, 1990-2013: A systematic analysis for the Global Burden of Disease Study 2013. Lancet 2015;385:117-71. https://doi.org/10.1016/S0140-6736(14)61682-2 . - DOI
-
- Pavin D, Delonca J, Siegenthaler M, et al. Long-term (10 years) prognostic value of a normal thallium-201 myocardial exercise scintigraphy in patients with coronary artery disease documented by angiography. Eur Heart J 1997;18:69-77. https://doi.org/10.1093/oxfordjournals.eurheartj.a015120 . - DOI - PubMed
-
- Takahashi Y, Miyagawa M, Nishiyama Y, et al. Performance of a semiconductor SPECT system: Comparison with a conventional Anger-type SPECT instrument. Ann Nucl Med 2013;27:11-6. https://doi.org/10.1007/s12149-012-0653-9 . - DOI - PubMed
-
- Songy B, Guernou M, Hivoux D, et al. Prognostic value of one millisievert exercise myocardial perfusion imaging in patients without known coronary artery disease. J Nucl Cardiol 2016;25:120-30. https://doi.org/10.1007/s12350-016-0601-5 . - DOI - PubMed
-
- Einstein AJ, Blankstein R, Andrews H, et al. Comparison of image quality, myocardial perfusion, and left ventricular function between standard imaging and single-injection ultra-low-dose imaging using a high-efficiency spect camera: The MILLISIEVERT study. J Nucl Med 2014;55:1430-7. https://doi.org/10.2967/jnumed.114.138222 . - DOI - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources