Phototherapy and the use of heat shields in very low birthweight infants
- PMID: 3389875
- PMCID: PMC1778903
- DOI: 10.1136/adc.63.5.552
Phototherapy and the use of heat shields in very low birthweight infants
Abstract
The irradiance produced by phototherapy units in use on a neonatal unit was measured. The phototherapy tubes were found to decline at a variable rate. Heat shields used to decrease heat and evaporative losses in the very low birthweight infant reduce irradiance and may appreciably limit the effectiveness of phototherapy.