Novel Technique for the Treatment of Type Ia Endoleak After Endovascular Abdominal Aortic Aneurysm Repair
- PMID: 33899573
- PMCID: PMC8276339
- DOI: 10.1177/15266028211010469
Novel Technique for the Treatment of Type Ia Endoleak After Endovascular Abdominal Aortic Aneurysm Repair
Abstract
Purpose: Open surgical repair of type Ia endoleak after endovascular aortic aneurysm repair/sealing (EVAR/EVAS) is associated with significant perioperative mortality and morbidity. Current endovascular redo techniques face limitations, especially when the infrarenal landing zone is inadequate and the previous endograft is rigid and features a short or no main body. We present a novel concept for the treatment of type Ia endoleak using a custom-made branched device.
Technique: The 5-branch-device (Cook Medical, Bloomington, IN, USA) consists of a nitinol skeleton with branches, covered with a low-profile polyester fabric loaded in an 18F sheath. The device features a minimum of 2 proximal sealing stents and includes branches for renovisceral vessels as well as an additional 8 mm branch for the contralateral iliac limb. Implantation and sealing in the renovisceral vessels is carried out in standard fashion, using transfemoral and transaxillary access. Distal sealing is achieved by tapering the branched component into the ipsilateral iliac limb and using a bridging balloon-expandable or self-expandable stent-graft through the additional branch to the preexisting contralateral iliac limb.
Conclusion: Treatment of type Ia endoleak with a new custom-made device enables sufficient proximal seal while minimizing suprarenal aortic coverage and facilitates adequate component overlap. The low profile branched design accommodates implantation through the preexisting endograft and catheterization of target vessels.
Keywords: EVAR; EVAS; additional branch; bEVAR; custom-made device; type Ia endoleak.
Conflict of interest statement
Figures
References
-
- Moulakakis KG, Dalainas I, Mylonas S, et al. Conversion to open repair after endografting for abdominal aortic aneurysm: a review of causes, incidence, results, and surgical techniques of reconstruction. J Endovasc Ther. 2010;17:694–702. - PubMed
-
- Ascoli Marchetti A, Oddi FM, Vacca F, et al. The safety of EVAS surgical conversion in a comparative monocentric analysis. Ann Vasc Surg. 2020;68:310–315. - PubMed
-
- Jordan WD, Jr, Mehta M, Varnagy D, et al. Results of the ANCHOR prospective, multicenter registry of EndoAnchors for type Ia endoleaks and endograft migration in patients with challenging anatomy. J Vasc Surg. 2014;60:885–892. - PubMed