Experimental Analysis of Pressure and Flow Alterations During and After Insertion of a Multilayer Flow Modulator into an AAA Model with Incorporated Branch

Affiliations

Experimental Analysis of Pressure and Flow Alterations During and After Insertion of a Multilayer Flow Modulator into an AAA Model with Incorporated Branch

Simon Tupin et al. Cardiovasc Intervent Radiol. 2021 Aug.

Abstract

Purpose: The multilayer flow modulator (MFM) device has been used for the treatment of abdominal aortic aneurysm (AAA) for over a decade. Although several clinical studies have been published, criticism and concern over the device efficacy remain, as no quantitative analysis that describes its mechanism has been performed yet. The aim of this study was to experimentally evaluate the effect of MFM device deployment on aneurysmal pressure and branch perfusion.

Materials and methods: An experimental flow and pressure monitoring system was developed to analyze the MFM deployment procedure performed by a qualified radiologist in AAA geometries with and without side branch. Particle image velocimetry experiments were then conducted on models with and without MFM device to evaluate and compare flow patterns and local flow velocity and vorticity in the aneurysm.

Results: The experiments revealed no significant change in pressure and flow rate during and after deployment of the MFM device. The flow rate of the incorporated branch was fully preserved. On both models, the aneurysmal flow velocity was significantly reduced. In addition, the device modified local flow patterns, reducing vorticity and better feeding the incorporated branch.

Conclusion: This experimental study provides the basis for a better understanding of the mechanism of the MFM device, which allows intra-aneurysmal flow to decrease while preserving incorporated branch flow and reducing the risk of type II endoleak. The experimental system developed for this study was effective in simulating an endovascular procedure and studying the safety and effectiveness of endovascular devices.

Keywords: Abdominal aortic aneurysm; Branch perfusion; Endovascular deployment; Intra-aneurysmal pressure; Multilayer flow modulator; Particle imaging velocimetry.

PubMed Disclaimer

Similar articles

References

    1. Sakalihasan N, Michel J-B, Katsargyris A, Kuivaniemi H, Defraigne J-O, Nchimi A, et al. Abdominal aortic aneurysms. Nat Rev Dis Prim Cham. 2018;4:34.
    1. Wanhainen A, Verzini F, Van Herzeele I, Allaire E, Bown M, Cohnert T, et al. Editor’s choice–European Society for Vascular Surgery (ESVS) clinical practice guidelines on the management of abdominal aorto-iliac artery aneurysms. Eur J Vasc Endovasc Surg. 2019;57:8–93. - PubMed
    1. Murray TÉ, Brennan P, Maingard JT, Chandra RV, Little DM, Brooks DM, et al. Treatment of visceral artery aneurysms using novel neurointerventional devices and techniques. J Vasc Interv Radiol. 2019;30:1407–17. - PubMed
    1. Sultan S, Hynes N, Kavanagh EP, Diethrich EB. How does the multilayer flow modulator work? The science behind the technical innovation. J Endovasc Ther. 2014;21:814–21. - PubMed
    1. Veith FJ, Baum RA, Ohki T, Amor M, Adiseshiah M, Blankensteijn JD, et al. Nature and significance of endoleaks and endotension: summary of opinions expressed at an international conference. J Vasc Surg. 2002;35:1029–35. - PubMed

LinkOut - more resources