Performance of Repeat BinaxNOW Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Testing in a Community Setting, Wisconsin, November 2020-December 2020
- PMID: 33909068
- PMCID: PMC8135465
- DOI: 10.1093/cid/ciab309
Performance of Repeat BinaxNOW Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Testing in a Community Setting, Wisconsin, November 2020-December 2020
Abstract
Repeating the BinaxNOW antigen test for severe acute respiratory syndrome coronavirus 2 using 2 groups of readers within 30 minutes resulted in high concordance (98.9%) in 2110 encounters. Same-day repeat antigen testing did not significantly improve test sensitivity (77.2% to 81.4%) while specificity remained high (99.6%).
Keywords: BinaxNOW; COVID-19; SARS-CoV-2; antigen testing; diagnostic performance.
Published by Oxford University Press for the Infectious Diseases Society of America 2021.