RF electrode-tissue coverage significantly influences steam pop incidence and lesion size
- PMID: 33928696
- DOI: 10.1111/jce.15063
RF electrode-tissue coverage significantly influences steam pop incidence and lesion size
Abstract
Background: Steam pops are a rare complication associated with radiofrequency (RF) ablation and are hard to predict. The aim of this study was to assess the influence of coverage between the RF ablation electrode and cardiac tissue on steam pop incidence and lesion size.
Methods and results: An ex vivo model using porcine cardiac preparations and contact force sensing catheters was designed to perform RF ablations at different coverage levels between the RF electrode and cardiac tissue. During coverage level I, only the distal part of the ablation electrode was in contact with tissue. During coverage level II half of the ablation electrode, and during coverage level III the entire ablation electrode was embedded in tissue. RF applications (n = 60) at different coverage levels I-III were systematically performed using the same standardized ablation protocol. Ablations during coverage level III resulted in a significantly higher rate of steam pops (100%) when compared to ablations during coverage level II (10%) and coverage level I (0%), log rank p < .001. Coverage level I ablations resulted in significantly smaller lesion depths, diameters, and impedance drops when compared to higher coverage level ablations, p < .001. In the controlled ex vivo model, there was no difference in applied contact force or energy between different coverage levels.
Conclusions: The level of coverage between RF electrode, cardiac tissue, and the surrounding fluid significantly influenced the incidence of steam pops in an ex vivo setup. Larger coverage between RF electrode and tissue resulted in significantly larger lesion dimensions.
Keywords: lesion size; radiofrequency ablation; steam pop.
© 2021 Wiley Periodicals LLC.
Comment in
-
Tissue coverage matters.J Cardiovasc Electrophysiol. 2021 Jun;32(6):1600-1601. doi: 10.1111/jce.15055. Epub 2021 May 12. J Cardiovasc Electrophysiol. 2021. PMID: 33928705 No abstract available.
References
-
- Neuzil P, Reddy VY, Kautzner J, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol. 2013;6:327-333.
-
- Kowalski M, Parikh V, Salcido JR, et al. Same-day discharge after cryoballoon ablation of atrial fibrillation: a multicenter experience. J Cardiovasc Electrophysiol. 2020.
-
- Kewcharoen J, Techorueangwiwat C, Kanitsoraphan C, et al. High-power short duration and low-power long duration in atrial fibrillation ablation: A meta-analysis [published online ahead of print November 18, 2020]. J Cardiovasc Electrophysiol. 32(1):71-82. https://doi.org/10.1111/jce.14806
-
- Creta A, Ventrella N, Providência R, et al. Same-day discharge following catheter ablation of atrial fibrillation: a safe and cost-effective approach. J Cardiovasc Electrophysiol. 2020;31(12):3097-3103.
-
- Yokoyama K, Nakagawa H, Shah DC, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol. 2008;1:354-362.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources