Case Reports
. 2021 May;34(3):252-256.
doi: 10.1111/sdi.12969. Epub 2021 May 3.

Hemodialysis catheter malpositioned into the mediastinum: Lessons from two cases

Affiliations
Case Reports

Hemodialysis catheter malpositioned into the mediastinum: Lessons from two cases

Seok Hui Kang et al. Semin Dial. 2021 May.

Abstract

Hemodialysis catheter (HDC) using a tunneled cuffed catheter (TCC) is a viable option for acute hemodialysis. Vessel injury during catheter insertion can lead to malpositioning of the HDC into the mediastinum. We herewith present two malpositioned HDCs in various clinical settings. For the first case, the HDC was inserted without fluoroscopic guidance, but in the second case, the HDC was inserted under fluoroscopic guidance. Our first case involved HDC insertion via the right internal jugular vein (IJV), and the second case involved HDC insertion via the left IJV. We clinically suspected HDC malposition because of poor aspiration, despite good infusion via the two lumens. Chest radiography, computed tomography, and venography showed various abnormal positions of the HDC. Our case report provides additional information about malpositioned or malfunctioning HDCs in various clinical settings.

Keywords: hemodialysis; internal jugular vein; mediastinum; tunneled cuffed catheter.

PubMed Disclaimer

References

REFERENCES

    1. Lok CE, Huber TS, Lee T, et al. KDOQI clinical practice guideline for vascular access: 2019 update. Am J Kidney Dis. 2020;75(Suppl 2):S1-S164.
    1. Niyyar VD, Work J. Interventional nephrology: core curriculum 2009. Am J Kidney Dis. 2009;54:169-182.
    1. Tong MK, Siu YP, Ng YY, Kwan TH, Au TC. Misplacement of a right internal jugular vein haemodialysis catheter into the mediastinum. Hong Kong Med J. 2004;10:135-138.
    1. Sainaresh VV, Jain SH, Patel HV, Shruti G, Trivedi HL. Malplacement of hemodialysis permcatheter into anterior mediastinum following innominate vein perforation. Semin Dial. 2012;25:95-96.
    1. Kosmadakis G, Correia Eda C, Carceles O, Somda F, Bros C, Aguilera D. Mediastinal malposition of a left jugular tunneled hemodialysis catheter with bilateral hemothorax and pneumo-hemo-mediastinum. Saudi J Kidney Dis Transpl. 2015;26:786-789.